9.12.7.16
:DGIO:SET:OUT
{x}
:LOW[?]
Description:
Set or query the lower limit value for the limit value check of each
output.
x represents outputs 1 to 4.
Only relevant for the selected compare operation.
Mode:
Command or query
Parameter:
{<Value>} within the range of -999999.999 .. +999999.999
The significance is determined by the selected measuring unit.
Measuring unit
Limit value unit
Example
Value
Tesla
mT
123.4 mT
123.4
A/m
A/m
1.2 kA/m
1200.0
Gauss
G
12.3 kG
12300.0
Oersted
Oe
12.3 kOe
12300.0
Response:
like parameter
*RST Value:
not relevant
Example:
send
:DGIO:SET:OUT2:LOW -1234.56<LF>
send
:DGIO:SET:OUT2:LOW?<LF>
receive -1234.56<CR><LF>
9.12.7.17
:DGIO:SET:OUT
{x}
:HIGH[?]
Description:
Set or query the upper limit value for the limit value check of each
output.
x represents outputs 1 to 4.
Only relevant for the selected compare operation.
Mode:
Command or query
Parameter:
{<Value>} within the range of -999999.999 .. +999999.999
The significance is determined by the selected measuring unit.
Measuring unit
Limit value unit
Example
Value
Tesla
mT
123.4 mT
123.4
A/m
A/m
1.2 kA/m
1200.0
Gauss
G
12.3 kG
12300.0
Oersted
Oe
12.3 kOe
12300.0
Response:
like parameter
*RST
Value:
not relevant
Example:
send
:DGIO:SET:OUT2:HIGH 1234.56<LF>
send
:DGIO:SET:OUT2:HIGH?<LF>
receive 1234.56<CR><LF>