Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 59 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.7.3
Operating Mode FLOW
In this operating mode, the current measuring value is transmitted every 100ms. The
measuring value is enclosed within <STX> and <ETX>. The syntax is:
<STX>snnnn.<ETX>
<STX>snnn.n<ETX>
<STX>snn.nn<ETX>
<STX>sn.nnn<ETX>
The date is transmitted with sign, four decimal digits and a point as decimal
separator. An overflow is diplayed as +9999.
9.8
Connector Cable EIA-232
For direct connection to a computer, use a null modem cable. In a null modem cable
the data transfer lines TxD and RxD are crossed. The signal ground wire GND is
pulled through.
Handshake lines (DTR/DSR or RTS/CTS) are not used.
Read further details in the instructions of your controller. If necessary, the control lines
(DTR/DSR or RTS/CTS) must be bridged in the connector of the external contoller.
The gaussmeter can be connected to a standard computer with a COM interface.
Figure 28 Null Modem Cable