Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 80 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.3
Peak Value Function
9.12.3.1
:PEAK:NULL
Description:
Reset the current peak values.
Mode:
Command
Parameter:
none
Response:
none
*RST
Value:
not relevant
Example:
send
:PEAK:MODE NULL<LF>
9.12.3.2
:PEAK:FAST:READ?
Description:
Query the largest value of minimum and maximum of the fast peak
value detection, according to its absolute value.
Mode:
Query
Parameter:
none
Response:
Measuring value
*RST
Value:
not relevant
Example:
send
:PEAK:FAST:READ?<LF>
receive -9.00<CR><LF>
9.12.3.3
:PEAK:FAST:READ:MIN?
Description:
Query the minimum value of the fast peak value detection.
Mode:
Query
Parameter:
none
Response:
Measuring value
*RST
Value:
not relevant
Example:
send
:PEAK:FAST:READ?<LF>
receive -9.00<CR><LF>
Trial Version
Subje
ct to chan
ge