Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 63 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.10.2.3
Command Separator “;“
Several commands within the same command line string are separated by a
semicolon. By means of a semicolon the indicated path is not changed. The two
following statements have the same meaning:
CMD:LOW 1;:CMD:HIGH 5
CMD:LOW 1;HIGH 5
9.10.2.4
Parameter Separator “,“
If you require several parameters in one command, they have to be separated by a
comma.
9.10.2.5
Use of Blanks
A parameter must be separated from a command keyword by blanks (tab or space).
Blanks are usually only ignored in parameter lists.
9.10.2.6
Query Commands
The controller can send out commands at any time, however a SCPI device (here the
gaussmeter) will only answer, if it has expressly been instructed to do so. Only query
commands (commands that end with a question mark) prompt the device to send a
response. For queries, the device displays either measuring values or internal device
settings.
Note
If you send two query commands without having read the reply to the
first and thereupon try to read the first response, you might receive
some data of the first reply, followed by the complete second
response. Therefore do not send any query command without having
read the reply beforehand. Commands and queries should not be
sent via the same program line. This might result in an overflow of
the raw data buffer, in the event that too many data are created.
9.10.2.7
Common Commands
Commands starting with an asterisk are called common commands. The commands
with asterisk are used for controlling status operations in the gaussmeter.