Gaussmeter IGM11 Operating Instructions
Chapter 8 Signal Interfaces
Page 42 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
8
Signal Interfaces
8.1
Introduction
The device has three digital inputs and four digital outputs.
The inputs and outputs are designed for a maximum voltage of 24 V. The inputs and
outputs are insulated and potential free. The inputs resistances and the maximum
allowed load are listed in the technical data. Both the inputs and the outputs can be
operated with both polarities (bipolar).
All input and output signals are on connector J1.
The function of each signal can be configured independently. See documentation of
the set-up menu.
The inputs and outputs each have a common connection.
The common potential of the inputs is on connector CI.
The common potential of the outputs is on connector CO.
The common potentials are on the power supply connector (J2); this simplifies the
wiring of the common potentials.
Note
The additional connectors PH1, PH2, REF, STR and GND of J1 are used
for specific features. Do not connect anything to these connectors!
Voltages higher than 24 V to these connectors might damage the
device.