68
BE 57-660 Eclipse Model 700 Guided Wave Radar Transmitter
Operational Mode
Current Consumption
Vmin
Vmax
HART
General Purpose
4mA
20mA
16.25V
11V
36V
36V
Intrinsically Safe
4mA
20mA
16.25V
11V
28.6V
28.6V
Fixed Current-Solar Power Operation (PV transmitter via HART)
General Purpose
10mA
¿
11V
36V
Intrinsically Safe
10mA
¿
11V
28.6V
HART Multi-Drop Mode (Fixed Current)
Standard
4mA
¿
16.25V
36V
Intrinsically Safe
4mA
¿
16.25V
28.6V
3.6.8 Power Supply Requirements
3.6.8.1 Safe Operating Area
¿
Start-up current 12 mA minimum.
3.6.8.2 Supply Voltage
0
16.25 V
591
Ω
24 V
36 V
Safe Operating Area
Vsupply
R
Loop
Digital Solar Mode
Typical HART
4 - 20 mA
Operating Area