47DR OPERATING MANUAL EM-9509-B Rev.7
3. SETTING INPUT TYPE
Set input type according to the RTD to use.
3.1 INPUT TYPE LIST
■
TEMPERATURE UNIT: °C
MAIN DISPLAY
FUNCTION
USABLE RANGE
CONFORMANCE RANGE
DEFAULT VALUE
TYPE 4
Pt 100 (JIS ’97, IEC)
-230.00 to +880.00
-200.00 to +850.00
TYPE 4
TYPE 5
Pt 50 Ω (JIS ’81)
-230.00 to +679.00
-200.00 to +649.00
TYPE 7
Pt 1000
-230.00 to +880.00
-200.00 to +850.00
TYPE 1
JPt 100 (JIS ’89)
-230.00 to +530.00
-200.00 to +500.00
TYPE 3
Pt 100 (JIS ’89)
-230.00 to +680.00
-200.00 to +650.00
■
TEMPERATURE UNIT: °F
MAIN DISPLAY
FUNCTION
USABLE RANGE
CONFORMANCE RANGE
DEFAULT VALUE
TYPE 4
Pt 100 (JIS ’97, IEC)
-382.0 to +1616.0
-328.0 to +1562.0
TYPE 4
TYPE 5
Pt 50 Ω (JIS ’81)
-382.0 to +1254.2
-328.0 to +1200.2
TYPE 7
Pt 1000
-382.0 to +1616.0
-328.0 to +1562.0
TYPE 1
JPt 100 (JIS ’89)
-382.0 to +986.0
-328.0 to +932.0
TYPE 3
Pt 100 (JIS ’89)
-382.0 to +1256.0
-328.0 to +1202.0
IMPORTANT
• The display accuracy is assured within the conformance range though indicating within the usable range is available.
• Indication less than ‘-200.00’ is not available when the temperature unit is set to °C and the decimal point position to 2
decimal places.
• The bargraph lower and upper limits, analog outputs 0% and 100% are returned to the default values per input type
when the input type has been changed. All alarm setpoints of the current bank No. are disabled (reset to ‘------’ status).
It is recommended to record the current settings as necessary.