68
47DR OPERATING MANUAL EM-9509-B Rev.7
8.2 BARGRAPH LOWER LIMIT/UPPER LIMIT
8.2.1 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the settings.
1
Hold down Init/
↑
button for 3 seconds or more to move on to
Initial Setting Mode.
• The input type is indicated.
• The SD indicates ‘RTTYPE’.
2
Press Alarm/
↓
or Init/
↑
button to go to the bargraph lower limit
setting.
• The bargraph lower limit is indicated.
• The SD indicates ‘BAR-L’.
3
Press Shift button to shift the display into the setting standby
mode. Then press Shift and Up buttons to set the bargraph
lower limit.
• The sixth digit starts blinking, to which you can apply changes.
• Set within the range of -999.99 to 9999.99.
NOTE
Set to 2 decimal places regardless of the decimal point position setting.
4
Press Alarm/
↓
button to apply the new value and go to the
bargraph upper limit setting.
• The bargraph lower limit is registered.
• The bargraph upper limit is indicated.
• The SD indicates ‘BAR-H’.
2,4,
6,7
1,2,
6,7
3,5
3,5
Blinking