8 Troubleshooting
This error is displayed if power supply is
interrupted during normal operation,
causing the bulb to extinguish. It is also
displayed if voltage is applied while lamp
is switched off or if the bulb or the igniter
is defect.
Make sure to provide stable power
supply.
Check the bulb and change if necessary.
Check if drawer is completely closed. Otherwise the internal power connection of the
lamp is not correctly closed.
Serious hardware problem. Send equipment to Henkel Technical Service for checking
and repair.
This error is displayed if
– Ignition was attempted although no bulb
has been installed,
– Normal bulb service life has been
exceeded considerably, or
– The bulb has overheated-
Install bulb.
Install new bulb.
Allow bulb to cool down (approx. 5 minutes), then re-ignite.
No voltage measured at bulb during
ignition.
– Cord required for power supply to lamp
housing is not connected.
Check connector cord (XS4) on lamp
housing and controller (XS4).
This error is displayed if
– Data cable for lamp housing not
connected.
Check cord on lamp housing (XS3) and
controller (XS3).