36
8 Troubleshooting
Insert the sensor cable plug into the
relevant socket on the PCB.
Re-assemble top cover and tighten the
4 screws.
CAUTION!
Be careful not to disturb the ground cable!
8.6 Sensor Cable Replacement
To disassemble lamp housing from the cure chamber, remove the 4 screws.
Lift lamp housing off.
Disconnect plug connection.
Disassemble base plate - remove the 10 screws.
Disconnect sensor cable from the connector on the sensor PCB and connect plug of
new cable.
Pass both sensor cables through one of the rear ventilation slots into the lamp housing.
Re-assemble base plate.
Re-connect plug connection.
Re-assemble lamp housing to the cure chamber and screw down.
Remove the 4 screws which hold down the red top cover.
Lift the top cover off.
CAUTION!
Be careful not to disturb the ground cable!
Disconnect sensor cable from the plug on the PCB and connect plug of new cable.
Re-assemble top cover and tighten the 4 screws.
8.7 Change of Reflector
Unlock and open the drawer.