6.4  Encoder 

The user must provide a suitable encoder.  The requirements are: 

  Differential quadrature output (maximum 18 V) 

  +5 VDC power supply input 

The encoder interface on the Master provides the following: 

  X4 quadrature decoding 

  Maximum 300 kHz count rate 

  +5 VDC encoder voltage output 

  Maximum 200 mA encoder current output 

  Input voltage range: -8 to 13V 

  Differential voltage: +/- 200mV 

We  recommend  using  RS422/RS485  encoders  instead  of  TTL  encoder  for  reliable 
performance.  Additionally,  the  user  should  choose  an  encoder  with  the  appropriate 
number of pulses per revolution to match the transport mechanism and speed. 

6.5  Wiring and Connections 

Refer  to  the 

FireSync  Network  User  Manual

  for  details  on  connecting  the  power 

supplies,  encoder,  Master,  Slave,  Station,  and  sensors  in  a  scan  3x00  sensor 
system. 

6.5.1 

Safety Interlock 

Safety  Interlock  is  a  signal  controlled  by  the  Safety  input  on  both  the  Master  and  the 
Slave.    Safety-  must  be  connected  to  the  48VGND  and    must  be  energized  by 
+48  VDC  before  any  lights  in  the  system  will  receive  power.    If  this  input  is  disabled, 
although the sensors are powered and running, the light sources are disabled. 

Systems with Safety Interlock require: 

  Safety interlock-enabled sensors 

  FireSync Sensor Cordsets Version 2 

  An  external  switch  connecting  the  Safety  Interlock  signal  at  the  Master  to  +48 

VDC 
 

Refer  to  the 

FireSync  Network  User  Manual

  for  details  on  wiring  the  Safety  Interlock 

circuit. 

Содержание chroma+scan 3155

Страница 1: ...chroma scan 3x00 Document Version 4 11 12 30...

Страница 2: ...r machine readable form without prior written consent of LMI Technologies Inc Trademarks and Restrictions DynaVision chroma scan Selcom FireSync and Sensors That See are registered trademarks of LMI T...

Страница 3: ...Environment 12 2 2 Temperature and Humidity 12 2 3 Maintenance 12 2 4 Laser Lifetime 12 3 Getting Started 13 3 1 Components 13 3 1 1 Sensors 14 3 1 2 Master 15 3 1 3 Cables 15 3 1 4 Power Supply 15 3...

Страница 4: ...6 6 Frame Design 39 6 6 1 chroma scan 3350 System Configuration 39 6 6 2 chroma scan 3155 3250 System Configuration 40 6 7 System Calibration Target 41 7 Software 42 7 1 FireSync Station 42 7 1 1 Inst...

Страница 5: ...7 4 1 Free Mode 78 7 4 2 Calibration Mode 79 7 4 3 Web Mode 80 7 4 4 Detection Mode 83 7 5 Health Indicators 85 7 6 Sample Code 85 7 7 Multiple Server Usage 85 7 8 Server Simulation 86 8 System Update...

Страница 6: ...t 1040 except for deviations pursuant to Laser Notice No 50 dated July 26 2001 Class 3B lllb sensors are sold only to qualified OEM s as components for incorporation into their own equipment The senso...

Страница 7: ...reflex the wavelength of light must be in the visible range 400 nm to 700 nm The Maximum Permissible Exposure MPE for visible radiation for 0 25 second is 25 Watt per square meter which is equivalent...

Страница 8: ...seful length Specular reflection Not required Prevent unintentional reflections Prevent unintentional reflections Eye protection Not required Not required Required under special conditions Laser safet...

Страница 9: ...pplying power to the lasers Beam Attenuators A permanently attached method of preventing human access to the laser radiation other than switches power connectors or key control must be employed On som...

Страница 10: ...to obtain the information booklet Regulations for the Administration and Enforcement of the Radiation Control for Health and Safety Act of 1968 HHS Publication FDA 88 8035 This publication containing...

Страница 11: ...ar washers on the mounting holes or through the earth ground connection located on the rear of the Master 1200 and 2400 FireSync Networking products should be installed inside electrical cabinets that...

Страница 12: ...to vibration or impact water oil or chemicals may splash onto the unit where static electricity is easily generated 2 2 Temperature and Humidity LMI 3D Sensors are rated for operation between 0 50 C...

Страница 13: ...nning This is followed by instructions on how to connect these components together and how to install and run our client software to acquire basic profile data to ensure sensor operation out of the bo...

Страница 14: ...14 3 1 1 Sensors chroma scan 3155 chroma scan 3250 chroma scan 3350...

Страница 15: ...Sensor network GigE switch Note Cable lengths for cs3x00 systems should not exceed 70m 3 1 4 Power Supply The user must provide a suitable 48 VDC power supply for the system The output from all power...

Страница 16: ...thernet ports Recommended Intel PRO 1000 Network Adaptor or equivalent x2 Pentium 4 Pentium D or Pentium Core Duo at 2 0 GHz or above Minimum 2 GB of RAM Minimum 10 GB of Hard Disk Space Windows XP SP...

Страница 17: ...a suitable Gigabit Ethernet Switch for the Sensor network The Gigabit Ethernet switch is connected to each Sensor through the Ethernet RJ45 connection through the Sensor s network cordset The Gigabit...

Страница 18: ...onnection properties window The client PC has now been configured to connect to a FireSync Station After this procedure is complete the station can be connected to via FireSync Client See Section 3 3...

Страница 19: ...ted devices are functioning normally For a detailed description of the FireSync Client application see the section entitled FireSync Client Complete in the Software chapter of this manual 3 3 1 Instal...

Страница 20: ...in the toolbar to display the Connect dialog To connect to a single server enter the IP address of the server to which you wish to connect and then click OK To connect to multiple servers simultaneous...

Страница 21: ...for each server and then click on the Health visualization tab as show below Health indicators can be used to help diagnose a wide variety of conditions Note that some indicators are updated constantl...

Страница 22: ...p is changed the number of sensor nodes in the device tree changes accordingly Assign a serial number to each sensor by right clicking on the sensor s node in the device tree and selecting a serial nu...

Страница 23: ...ensor click on the device tree node for each sensor and review the Sensor Health visualization tab As with the server health indicators some sensor health indicators are updated constantly while other...

Страница 24: ...ice tree and then select the Obstructions tab Press the Start button in the FireSync Client toolbar The Profile visualization tab will display live laser profile measurements This display can be used...

Страница 25: ...ld proven FireSync platform which provides a synchronized scalable distributed vision processing architecture for building reliable high performance systems High reliability and simple rapid installat...

Страница 26: ...and reliable method of capturing features on the top and bottom surfaces of the boards along the length of the system 4 5 Tracheid chroma scan 3250 sensors have the ability to measure the tracheid cel...

Страница 27: ...e models and have safety interlock for lasers and LED control Safety interlocked sensors require the use of the FireSync Sensor Cordset Version 2 to provide a 48 VDC control signal to each sensor in t...

Страница 28: ...olution 0 127 mm 0 005 in Profile Y Resolution 1m s 197 fpm 1 25m s 246 fpm for chroma scan 3250 0 5 mm 0 020 in 5 3 Specifications Operating Temperature Non condensing 0 50 C 32 122 F Input Power 48V...

Страница 29: ...29 5 4 Scan Zone 5 4 1 chroma scan 3350 Scan Zone Standoff SO 559 mm 22 0 in Measurement Range MR 152 mm 6 0 in Field of View FOV through entire range 610 mm 24 0 in...

Страница 30: ...30 5 4 2 chroma scan 3155 3250 Scan Zone Standoff SO 508 mm 20 0 in Measurement Range MR 203 2 mm 8 0 in Field of View FOV through entire range 609 6 mm 24 0 in...

Страница 31: ...31 5 5 Dimensions and Mounting 5 5 1 chroma scan 3155 Sensor Dimensions...

Страница 32: ...32 5 5 2 chroma scan 3250 Sensor Dimensions...

Страница 33: ...33 5 5 3 chroma scan 3350 Sensor Dimensions...

Страница 34: ...reflections in the system A more critical alignment can be made for aesthetic purposes The profile cameras are located behind the two red tinted windows and the vision cameras chroma scan 3350 sensor...

Страница 35: ...cal constraints of the chain design prevent mounting sensors there The Master is the central hub of the system and provides power encoder Safety Interlock and digital I O routing Each sensor and the e...

Страница 36: ...vision output from chroma scan 3350 sensors can exceed the networking and data processing capacity of a single station computer There should typically be one dedicated station for every 5 10 chroma sc...

Страница 37: ...x Contact QUINT 10 Amp power supply for 48 VDC The Phoenix QUINT series power supplies are DIN rail mounted devices that can be connected in parallel to increase the overall available power 48 VDC Mod...

Страница 38: ...iring and Connections Refer to the FireSync Network User Manual for details on connecting the power supplies encoder Master Slave Station and sensors in a chroma scan 3x00 sensor system 6 5 1 Safety I...

Страница 39: ...rs location and orientation is required to ensure that the sensors do not experience crosstalk and also to ensure that the board thickness is measured differentially The scan frame must also be connec...

Страница 40: ...rovides an additional 2 inches of coverage from the top sensors above this overlapped zone as well as an additional 2 inches below the overlapped zone from the bottom sensors An alternative system con...

Страница 41: ...he system The target illustrated below is designed for a chroma scan 3350 system with sensors spaced on 24 centers Note that the reference holes are spaced at 12 inches to provide a reference location...

Страница 42: ...ent user interface settings file format message formats and health indicators 7 1 FireSync Station Any PC meeting the requirements described in Section 3 1 5 can become a FireSync Station node by inst...

Страница 43: ...llation location This can be any valid location that does not contain a space in its path Start menu shortcuts are added as a convenience and can be renamed safely Note that a shortcut will be added t...

Страница 44: ...hat future upgrades loaded through the FireSync Host Protocol can proceed unimpeded 7 1 2 PC with Existing Station Software If the PC contains existing station software installed by other means the sh...

Страница 45: ...AC on Windows 7 FireSync Station software requires that UAC be disabled on Windows 7 machines To Disable the UAC follow these steps Browse to the User Accounts section in control panel Then ensure tha...

Страница 46: ...demonstrate or diagnose problems with a chroma scan 3x00 system The following sections describe how to use FireSync Client to configure a chroma scan 3x00 system 7 2 1 Installation The FireSync Clien...

Страница 47: ...dialog To connect to a single server enter the IP address of the server to which you wish to connect and then click OK To connect to multiple servers simultaneously click the Advanced button add the...

Страница 48: ...ngs can be specified using the System node in the FireSync Client device tree If multiple stations are connected FireSync Client will ensure the settings in the System node are sent to each connected...

Страница 49: ...ommended that once the system setup is complete and the system is operating as desired that the user creates a backup of all stations in the system Select each server one at a time click on the Server...

Страница 50: ...ion of the top mounted sensors Facing Toward or Facing Away The sensor logo and labels are visible when sensors are facing toward Y Orientation bottom Specifies the y orientation of the bottom mounted...

Страница 51: ...der resolution For example if the chain might jitter by 0 9 when stopped the corresponding number of ticks can be calculated from the encoder resolution A laser activation value of 0 disables the lase...

Страница 52: ...52 Recommended values 20 to 20 default is 0...

Страница 53: ...X Res X resolution length axis of profile data in mils per pixel Profile Y Res Y resolution width axis of profile outputs in mils per pixel Vision X Res X resolution length axis of vision outputs in m...

Страница 54: ...terpolation between two valid laser spots when resampling data within a single camera Overlap Blend Maximum amount of redundant camera information that can be blended together at the x boundary betwee...

Страница 55: ...detection trigger Global or Local An object is detected when the amount of visible target material measured along the x axis exceeds the Trigger Length threshold With a local trigger the visible targ...

Страница 56: ...ls excluding edge margins When objects exceed this width they will be segmented into multiple output messages Note the true maximum width of a transmitted output includes the maximum width and edge ma...

Страница 57: ...locator holes are not visible e g blocked by conveyor This feature can also be disabled if the x center settings provided for each sensor are known to be accurate Enable this field to manually remove...

Страница 58: ...files from the station after calibration diagnostics are complete When selected the Calibration Tab enables Calibration mode the Start Pause and Stop buttons can be used to run the system Calibration...

Страница 59: ...ick on the Details tab to review warnings and errors Another set of visualization tabs display scans of the calibration bar with the calibration results applied to the data There is one visualization...

Страница 60: ...er Setup Tab The Server Setup tab shown below contains settings for each chroma scan 3x00 server Top Sensors Count of top mounted sensors connected to this server Bottom Sensors Count of bottom mounte...

Страница 61: ...lient device tree health indicators are displayed in a visualization tab as shown below Health indicators can be used to help diagnose a wide variety of conditions Note that some indicators are update...

Страница 62: ...ined vision cameras and lights if present will be disabled and if no tracheid sections are defined tracheid cameras will be disabled If no sections of any kind are defined profile cameras and lasers w...

Страница 63: ...be greater than X0 Channel Numeric identifier of the TCP output channel to which this section will be routed For more information on configuring TCP IP data channels see the FireSync Host Protocol Us...

Страница 64: ...he settings described in the following sections 7 2 7 1 Sensor Setup Tab X Center Specifies the center position of the sensor measured along the x axis from the system zero reference left or right end...

Страница 65: ...ded from measurement Note Obstructions are used in Web mode and Detection mode but are not used in Calibration mode Profile points within obstructions are always excluded from the data used for detect...

Страница 66: ...lient device tree health indicators are displayed in a visualization tab as shown below Health indicators can be used to help diagnose a wide variety of conditions Note that some indicators are update...

Страница 67: ...order to use the event channel ensure that the client PC has a network interface on the 90 X X X subnet and that the Master s Link port is connected to the network To open an event channel connection...

Страница 68: ...VisionExposure BottomVisionExposure 100 BottomVisionExposure LinkCapacity 1000 LinkCapacity Setup Sampling ProfileXResolution 200 ProfileXResolution ProfileYResolution 20 ProfileYResolution VisionXRes...

Страница 69: ...Channel Section Sections Members Sensor Name 0 Name SerialNumber 1471 SerialNumber Enabled 1 Enabled XCentre 12000 XCentre Obstructions Sensor Sensor Name 1 Name SerialNumber 6755 SerialNumber Enable...

Страница 70: ...itive if the encoder count increases when the board travels in the forward direction 4x quadrature decoding is assumed EncoderMsgVersion 0 or 1 1 by default It is the version of the Free Mode Profile...

Страница 71: ...p mounted sensors Facing Toward 1 or Facing Away 0 The sensor logo and labels are visible when sensors are facing toward BottomYOrientation Specifies the y orientation of the bottom mounted sensors Fa...

Страница 72: ...the tracheid data that is sent to the client in Detection mode The following values are supported 200 250 500 and 1000 TracheidYResolution The y resolution mils per tracheid value across board width...

Страница 73: ...each detected object MaximumWidth The maximum width of a detection output mils including edge margins When objects exceed this width they will be segmented into multiple output messages EdgeFilter Sho...

Страница 74: ...gions are specified via sensor obstructions see Section 7 2 7 2 and typically represent conveyor hardware regions When this field is disabled the calibration will automatically detect conveyor hardwar...

Страница 75: ...ects that regularly appear within the field of view of the sensors such as chain runners the zones in which those objects appear must be identified before the system will operate correctly in Calibrat...

Страница 76: ...ommands The system calibration file may not exist if system calibration has not been successfully completed Deleting this file from the FireSync Station clears the current system calibration The follo...

Страница 77: ...characters e g FFAA and represents a y correction in mils The number of array elements is variable minimum 2 elements actual y offsets will be determined by resampling these y offsets at the required...

Страница 78: ...geometric measurements captured using lasers and profile cameras There are no guarantees regarding the order in which data from separate sensors will arrive at the client To toggle between Free Mode P...

Страница 79: ...le named Calibration xml that can be accessed using the FireSync Host Protocol Read File and Write File commands The format of this file is described in Section 7 3 3 Calibration Output Note In multi...

Страница 80: ...ntensity Too High Low 2007 reserved 2 k64s Reserved for internal use 7 4 3 Web Mode In Web mode the server applies calibration transformations and resamples data to the requested output resolutions bu...

Страница 81: ...erved 4 k64s Reserved for internal use height k64s Profile array height width k64s Profile array width reserved 2 k64s Reserved for internal use inputs height k64s Digital input bits one set per profi...

Страница 82: ...height k64s Digital input bits one set per tracheid row angle height width k8u Tracheid laser spot angles 0 179 scatter height width k8u Scatter ratio of the minor axis to the major axis of the best f...

Страница 83: ...value This message contains batched encoder and timestamp readings at the profile speed for accurate encoder diagnostics The batch size is set through the server settings file see Section 7 3 1 To tog...

Страница 84: ...use deviceId k64s Server serial number groupId k64s Group identifier Top 0 or Bottom 1 sectionId k64s Section identifier from XML Section Id sequenceIndex k64s Object Id increases with each unique ob...

Страница 85: ...d FireSync Health Data Channel Note currently there are no custom indicators for chroma scan 3x00 sensors or servers Consult the FireSync Host Protocol User s Manual for the complete lists of standard...

Страница 86: ...ny of the servers reading the start delay for each server determining the maximum start delay and adding this value to the current system time Start Time System Time Max Server 1 start delay Server N...

Страница 87: ...and The instructions below describe how to get started using the Server Simulator Installation Install c s 3x00 client package Navigate to the program folder C chromascan CS3x00 Client bin win32 Serve...

Страница 88: ...d with a data file When the playback engine is running each channel transmits its own data file Output Style The FireSync server is capable of sending data in multiple methods Unless otherwise specifi...

Страница 89: ...a channel socket 0 Data channel id 127 0 0 1 Client IP address 4500 Client port 6 Configure server to transmit to client channel 0 Data channel id 127 0 0 1 Client IP address 4500 Client port 11 Start...

Страница 90: ...ckup in the Server menu Choose an appropriate location to save the backup dat Repeat this step for each server in the system Note that the backup file is unique to each server in the system 2 Install...

Страница 91: ...lease software updates through the Support website These updates consist of a FireSync Client installer and b server upgrade for each model type The following procedure can be used to update the syste...

Страница 92: ...nt is replaced The following steps describe this procedure 7 Mount and connect all new components Sensors Masters Slaves Stations 8 Download the latest Client and Firmware files from http www lmi3d co...

Страница 93: ...ystem functionality and report any issues 21 Now the system should be in the ready state for customer application to run If the current installed Client is the same as the latest release steps 9 and 1...

Страница 94: ...ogies Inc warranty covers parts labor and the return shipping charges If the warranty stickers on the sensors are removed or appear to be tampered with LMI will void the warranty of the sensor 10 2 Re...

Страница 95: ...distributor or LMI directly Visit our website at www lmi3D com for the agent nearest you For more information on Safety and Laser classifications contact U S Food and Drug Administration Center for D...

Отзывы: