48
2. Select the 'Server' node in the device tree and from the 'Server' menu, select
‘Upgrade...’
3. Browse for the desired Firmware upgrade package (.dat) when prompted, then click
'OK' to begin the upgrade process. Ensure that you select the upgrade file specific to
the sensor type (31x0, 32x0, or 33x0) in your system.
4. Wait for the upgrade process to complete. This process requires several minutes.
5. After upgrading each Station in your system individually, please power-cycle all
components (sensors, masters, slaves, stations).
6. After power-cycling all components you can then connect to all stations
simultaneously.
Follow this procedure whenever a firmware update is released by LMI.
7.2.4
System Setup
After upgrading, the next step is to set up the system. Most settings can be specified
using the
System
node in the FireSync Client device tree. If multiple stations are
connected, FireSync Client will ensure the settings in the System node are sent to each
connected station.
Note:
When settings appear disabled in the FireSync Client UI, it can be for one of the
following reasons. First, it is not possible to edit settings while the system is running
–
press the Stop button to stop the system before attempting to modify settings. Second,
some settings are not available with particular sensor models. For example, tracheid
resolutions are not applicable to systems that are not equipped with tracheid cameras.
Finally, if the system has been calibrated, it will not be possible to edit many of the
system settings until the calibration is cleared. In this case, navigate to the
System
Calibration
tab and click the
Clear
button to remove the calibration. After the calibration
is removed, most settings will be editable again.
Содержание chroma+scan 3155
Страница 1: ...chroma scan 3x00 Document Version 4 11 12 30...
Страница 14: ...14 3 1 1 Sensors chroma scan 3155 chroma scan 3250 chroma scan 3350...
Страница 31: ...31 5 5 Dimensions and Mounting 5 5 1 chroma scan 3155 Sensor Dimensions...
Страница 32: ...32 5 5 2 chroma scan 3250 Sensor Dimensions...
Страница 33: ...33 5 5 3 chroma scan 3350 Sensor Dimensions...
Страница 52: ...52 Recommended values 20 to 20 default is 0...