![LM-Dental LM-ProPower CombiLED Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/lm-dental/lm-propower-combiled/lm-propower-combiled_user-manual_1932118008.webp)
Safety instructions
8
2.6. EMC Guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration - electromagnetic emissions
The LM-ProPower is intended for use in the electromagnetic environment specified below.
The customer or the user of the LM-ProPower should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The LM-ProPower uses RF energy only for its internal function.
Therefore, it’s RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The LM-ProPower is suitable for use in all establishments, in-
cluding domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration - electromagnetic immunity
The LM-ProPower is intended for use in the electromagnetic environment specified below. The customer or the user of the
LM-ProPower should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
±6 kV contact
±6 kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
IEC 61000-4-2
±8 kV air
±8 kV air
Electrical fast transient/
burst
±2 kV for power supply
lines
±2 kV for power sup-
ply lines
Mains power quality should be that
of a typical commercial or hospital
environment.
IEC 61000-4-4
±1 kV for input/output lines
±1 kV for input/output
lines
Surge
±1 kV differential mode
±1 kV differential
mode
Mains power quality should be that
of a typical commercial or hospital
environment.
IEC 61000-4-5
±2 kV common mode
±2 kV common mode
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be that of
a typical commercial or hospital en-
vironment. If the user of the LM-Pro-
Power requires continued operation
during power mains interruption, it is
recommended that the LM-ProPower
be powered from an uninterruptible
power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE U
T
is the AC mains voltage prior to application of the test level.
Содержание LM-ProPower CombiLED