4.2. How to connect the unit to the air supply
NOTICE!
Use only dry and filtered compressed air.
NOTICE!
Special skills required. Consult an approved technician to connect the unit to
the air supply.
1. Set the air supply OFF.
2. Verify that the air pressure obeys to the data in the
technical data section. See instructions “9. Technical
data” on page 39
3. Use only dry and filtered compressed air.
4. Push the hose onto the quick connect coupling.
5. Connect the other end of the hose to the air supply.
6. Set the air supply ON.
7. To disconnect, push orange ring and pull tubing out
when orange ring still is pushed in.
4.3. How to connect the unit for tap water (optional)
NOTICE!
Special skills required. Consult an approved technician to connect the unit to
the water supply.
1. Set the water supply OFF.
2. Verify that the water pressure obeys to the data
in the Technical data section. See instructions “9.
Technical data” on page 39
3. Verify that the water supply fulfils the medical
demands of hygiene.
4. Push the hose onto the quick connect coupling.
5. Connect the other end of the hose to the water
supply.
6. Set the water supply ON.
7. To disconnect, push orange ring and pull tubing out
when orange ring still is pushed in.