Troubleshooting CombiLED
37
6. Use the LM-ProPower Fixer to examine that the water hose inside the handpiece has not stuck in the
autoclave. Put the Fixer carefully into the handpiece from the hose connector side. (To prevent the hose
getting stuck, the water adjustment on the handpiece should be set to fully open before sterilization in
autoclave.)
7. If the problem stays – contact your dealer for support and indicate Error Code E-S04.30.
8.8. Tip does not fit smoothly onto the handpiece
1. Clean handpiece threads with compressed air and try with a new tip.
2. If the problem stays, the threads of the handpiece can be damaged and it is necessary to replace the
handpiece. Contact your dealer and indicate Error Code E-S06.
8.9. Handpiece is overheated during use
1. Examine that the fluid flow is sufficient (minimum 20 ml/min).
2. If the problem stays, replace the handpiece and contact your dealer.
8.10. Control panel does not respond when pressing the P-key
1. Check that the polisher tubing is connected properly i.e. the powder container cap is pressed in all
the way onto the connector. Check that the O-rings on the connector are in good condition. Replace
O-rings if worn. O-rings can be lubricated with a glycerine based lubricant.
2. If you have an additional polisher tubing, try replacing the polisher tubing.
3. If the problem still remains – contact your dealer for support and indicate Error Code E-X02.
8.11. Not sufficient or no water flow (Polisher)
1. Make sure that the Air Blow mode is not activated i.e. that the blue light adjacent to the crossed-out
water drop is not lit. If it is lit, press the P-key.
2. Adjust water control on powder container cap to fully open.
3. If the unit are equipped with a water bottle: Make sure that the medicament/water bottle is correctly
connected i.e. the cap is tightened and pushed in all the way onto the connector. Examine that the
O-rings on cap and connector are in good condition. Replace O-rings if worn. O-rings can be lubricated
with water soluble lubricant.
4. If the unit are equipped with a tap water connection: Make sure that the tap water hose is correctly
connected to the back of the device and that the tap water system is OK (tap/valve is open and possible
filter is OK).
5. Make sure that the nozzle is fully tightened.
6. If you have an additional nozzle, replace the nozzle.
7. If the problem stays – contact your dealer for support and indicate Error Code E-P04.