Content of delivery

14

3.6.  Medicament bottle (optional)

LM-ProPower has a medicament dispenser system, making the device independent of a fixed water supply 
connection. The medicament bottle can be used for either medicament solutions or ordinary clean water.  
Approved medicament solutions:

•  Clean water
•  Cetylpyridinium chloride
•  Chlorhexidine 
•  Essential oils
•  Hydrogen peroxide, 3% USP
•  Povidine Iodine, 10% solution
•  Saline solution
•  Sangurinara extract
•  Sodium hypochlorite 1% solution

The unit contains an electrically driven air compressor. When operating, the compressed air forces the fluid from 
the bottle through the hose and to the handpiece and to the tip/nozzle.

1

2

3

1

Medicament bottle

2

Bottle connector

3

Depressurisation button

3.7.  AirLED Polisher

1

2

4

3

1

O-ring (installed in cap)

2

AirLED tubing with handpiece

3

Water flow control

4

Powder container

Содержание LM-ProPower CombiLED

Страница 1: ...LMProPower CombiLED User Manual ...

Страница 2: ...ments Oy PL 88 Norrbyn rantatie 8 FI 21601 Parainen Finland Telephone 358 2 4546 400 Fax 358 2 4546 444 E mail info lm dental com Internet www lm dental com Copyright Copyright 2015 LM Instruments Oy All rights reserved The contents of this manual may be changed without notice No part of this manual may be reproduced in any form or by any means without permission in writing from LM Instruments Oy ...

Страница 3: ...ect the unit to the supply current 19 5 Operating instructions UltraLED Scaler 20 5 1 How to install UltraLED Scaler to use 20 5 2 Foot control functions for Scaler 22 5 3 Power and working modes for Scaler 22 5 4 How to operate with the UltraLED Scaler 23 5 5 After Scaler treatment 24 6 Operating instructions AirLED Polisher 25 6 1 How to install AirLED Polisher to use 25 6 2 Foot control functio...

Страница 4: ... 4 10 Warranty 40 10 1 Warranty terms 40 10 2 Delivery Information 41 10 3 Installation check list 42 10 4 Operational test 43 ...

Страница 5: ...essary to install the product and put it into servicing referred to the EMC data in chapter 2 6 EMC Guidance and manufacturer s declaration on page 8 Portable and mobile RF Radio Frequency communications equipment can have an unwanted effect on the product Connect the product to electricity water and compressed air supply which obeys the requirements in the Technical data Servicing Only approved s...

Страница 6: ...ed could cause death or injury CAUTION CAUTION indicates a dangerous situation that if not prevented could cause minor or moderate injury NOTICE NOTICE indicates a situation that if not prevented could result in material damage This general hazard symbol identifies important safety messages in this manual Carefully read understand and obey the messages When you see this symbol be alert your and yo...

Страница 7: ...any other procedures are strictly prohibited Before you do any maintenance work read and understand maintenance instructions 2 4 Intended and prohibited use LM ProPower CombiLED is designed for removal of biofilm and calculus on teeth cleaning discoloured teeth and other dental work where the ultrasonic vibration and or polishing is beneficial The unit must only be used by licensed dental and prof...

Страница 8: ...ESD 6 kV contact 6 kV contact Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 IEC 61000 4 2 8 kV air 8 kV air Electrical fast transient burst 2 kV for power supply lines 2 kV for power sup ply lines Mains power quality should be that of a typical commercial or hospital environment IEC 61000 4 4 1 kV for input ...

Страница 9: ...xed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and ...

Страница 10: ...tter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the fre...

Страница 11: ...12 LM ProPower fixer 6b 500 ml medicament bottle depending of the type of the device 3 2 General description LM ProPower CombiLED combines an effective piezoelectric UltraLED scaler and an AirLED Polisher in one versatile and ergonomic appliance The device s LED lights advanced electronics quality and high durability LM DuraGradeMAX tips enhance the execution of procedures which require great prec...

Страница 12: ...CombiLED unit 7 Powder container 2 Depressurisation button 8 Polisher handpiece incl tubing with an ErgoGrip and a nozzle mounted 3 Medicament bottle 9 Control panel 4 Scaler Handpiece with an ErgoGrip and a tip installed 5 Scaler water flow control ring 6 Foot control unit ...

Страница 13: ...6 5 1 Cleaning key 2 Working mode 3 key 3 Working mode 2 key 4 Working mode 1 key 5 Polishing mode indicator 6 Polishing mode key 7 Standby indicator 8 Scaling mode key 9 Scaling mode indicator 10 Irrigation mode indicator Scaler Water Jet mode indicator Polisher 11 Working mode 1 indicator 12 Dry mode indicator Scaler Air Blow mode indicator Polisher 13 Working mode 2 indicator 14 Working mode 3 ...

Страница 14: ...hlorhexidine Essential oils Hydrogen peroxide 3 USP Povidine Iodine 10 solution Saline solution Sangurinara extract Sodium hypochlorite 1 solution The unit contains an electrically driven air compressor When operating the compressed air forces the fluid from the bottle through the hose and to the handpiece and to the tip nozzle 1 2 3 1 Medicament bottle 2 Bottle connector 3 Depressurisation button...

Страница 15: ...Content of delivery 15 3 8 Foot Control unit 3 2 1 1 Foot control unit body 2 Foot control pedal 3 Connector for foot control cable ...

Страница 16: ...h the Safety Standards ANSI AAMI ES 60601 1 and CAN CSA C22 2 No 60601 1 08 Caution Consult accompanying documents Compliance label indicating compliance with the Medical Device Directive 93 42 EEC 0537 is the ID number of the Notified Body VTT Withstands autoclave temperature 135 C 275 F Type B applied part according to the degree of protection against electrical shock Fuse Input Output Please do...

Страница 17: ...at source Excessive heat can damage the electronics NOTICE Product is not suitable for use in the presence of flammable gases 4 1 How to install the unit Put the unit horizontally with the handpieces in the holders and the hose hanging free Put the unit where air is free to circulate on all sides and below it Do not put the unit on a wall or adjacent to heat source Do not put the unit in the immed...

Страница 18: ...upply 6 Set the air supply ON 7 To disconnect push orange ring and pull tubing out when orange ring still is pushed in 4 3 How to connect the unit for tap water optional NOTICE Special skills required Consult an approved technician to connect the unit to the water supply 1 Set the water supply OFF 2 Verify that the water pressure obeys to the data in the Technical data section See instructions 9 T...

Страница 19: ... plug must be classified as Hospital Grade 1 Verify that the voltage rating shown above type plate match the voltage of the AC power outlet 2 Verify that the AC power supply is supplied with the protective ground 3 Connect the power cable to the unit and the AC power supply All indicator lamps illuminates for a short period when the unit does a self check 4 The unit is standby when the green lamp ...

Страница 20: ...ent Notice that ultrasonic instruments with small diameters are subject to breakage at any time If not used correctly or with too much power or force the instrument will break CAUTION PERSONAL INJURY HAZARD Do not use nickel titanium root canal files with file holder Nickel titanium files breaks easily at high frequencies and may cause injury to the patient or to the user 5 1 How to install UltraL...

Страница 21: ...ators adjacent to the working modes are flashing to remind that you must select a working mode If some blue indicators are flashing alternatively make sure that the scaler handpiece is connected If the problem stays refer to the trouble shooting section See the recommended working mode that is identified on the tip E and push the related working mode key F on the keyboard Remember to choose right ...

Страница 22: ... foot control from 0 to 100 in each working mode See the recommended working mode that is identified on the tip Press the related working mode key on the keyboard Dry mode for scaling Press one time of the S key to activate the dry mode for scaling without water medicament Irrigation mode Press two times of the S key to activate the irrigation mode You can also activate the irrigation mode in the ...

Страница 23: ... first 2 mm of the tip are the most efficient Scaling rarely requires more than 50 power level of each working mode Hard to remove calculus might require a higher power setting The side of the tip end will be applied to the tooth so that the movement of the tip is parallel to the tooth surface Make sure that contact between the tip and the tooth surface is kept during scaling Keep moving the tip s...

Страница 24: ...cleaning function See instructions 7 2 Automatic cleaning function for the scaler and polisher on page 30 3 After the cleaning cycle is finished turn off the tip counterclockwise with torque wrench B 4 Squeeze the ErgoGrip carefully at the top and at the same time move it off the handpiece C Do not squeeze too hard at the ErgoGrip as this make the removal difficult 5 Press the depressurisation but...

Страница 25: ...er air polishing powders 6 1 How to install AirLED Polisher to use PREPARATIONS BOTTLE VERSION 1 Fill the medicament bottle with water or with approved medicament solution Refer to the approved medicament solutions See instructions 3 8 Medicament bottle optional on page 14 2 Turn the bottle cap onto the bottle and push it onto the connector 3 Make sure that the unit is connected to the air supply ...

Страница 26: ...ndicator lamp illuminates 8 The indicators adjacent to the working modes keys are flashing to remind that you must select a working mode If some blue indicators are flashing alternatively make sure that the polisher is connected If the problem stays refer to the trouble shooting section Choose working mode 2 or 3 by pressing the related working mode key on the keyboard 9 Adjust water flow for opti...

Страница 27: ...g mode 2 A Water Jet cleaning The Water Jet cleaning mode is cleaning without powder Select the Water Jet cleaning mode with the foot control pressed down in the leftmost position Only water and air comes out from the nozzle User can also select the Water Jet cleaning mode by pressing P key three times on the control panel Air Blow mode Press one time of the P key to activate the Air blow mode for...

Страница 28: ...D SODIUM BICARBONATE Do not perform Sodium Bicarbonate polishing treatment for patients who have a renal insufficiency are on sodium restricted diet have allergic reactions to flavour containing powders Use neutral powder instead are on a long term steroid or diuretic therapy Do not direct the jet powder towards filling crowns or bridgework as this could damage these restorations Do not direct the...

Страница 29: ...ning is sufficient to remove the biofilm from the pocket 5 Set OFF the device before lifting out the nozzle from the pocket 6 6 After Polisher treatment NOTICE Air purge the polisher handpiece after each treatment to prevent clogging NOTICE Immediately after using any type of medicament in the medicament bottle run the automatic cleaning cycle with clean water in the medicament bottle for both the...

Страница 30: ...e active a press on the C key will air purge the handpiece for some seconds Air purge the handpiece after each treatment to prevent clogging 7 2 Automatic cleaning function for the scaler and polisher NOTICE The automatic cleaning function must always be performed with the polisher nozzle installed in the handpiece 1 Set OFF the scaler or polisher by pressing the S or P key for 3 seconds The blue ...

Страница 31: ...ler handpiece NOTICE Do not sterilize any scaler or polisher accessories using dry heat or chemical autoclaves This may damage the material Clean with a soft cloth and use a surface disinfectant suitable for hard plastics Cover and control panel Polisher tubing with handpiece Clean at max 65 C Powder container Medicament bottle Cap for the medicament bottle ...

Страница 32: ...ece after each treatment See instructions 6 4 Air purge quick cleaning function on page 27 Run the automatic cleaning cycle with clean water for both the scaler and polisher See instructions 7 2 Automatic cleaning function for the scaler and polisher on page 30 Clean off the cover control panel handpieces and the hoses with a soft cloth Use a surface disinfectant suitable for hard plastics Clean t...

Страница 33: ... 1 and cap with dry compressed pressurized air 1 2 4 3 Weekly bottle version Run the automatic cleaning cycle with an anti microbial cleaning agent solution in the bottle for both the scaler and polisher See instructions 7 2 Automatic cleaning function for the scaler and polisher on page 30 We recommend to use a separate bottle for the cleaning agent solution Concerning exposure times of cleaning ...

Страница 34: ...re than 2 mm Prolonged use could cause tip breakage and injury to the patient Notice that ultrasonic instruments with small diameters are subject to breakage at any time If not used correctly or with too much power or force the instrument will break Do an inspection of the tips When a tip is bent altered or worn more than 2 mm it will loose performance and you must interchange the tip Examine the ...

Страница 35: ...hat the foot control cable is connected at both ends and is not damaged The blue lights adjacent to the S and P keys are flashing alternately when pressing the foot control The Scaler Polisher has not been switched on Press the S key to select the scaler or press the P key to select the polisher The blue light adjacent to the S key is flash ing alternately with three other blue lights on the contr...

Страница 36: ...he problem stays contact your dealer for support and indicate Error Code E S01 No vibration or E S02 Low power vibration 8 6 Weak tip vibration 1 Make sure that the tip is tightly tightened We recommend that you use a torque wrench and tighten clockwise until the torque wrench slides The torque wrench prevents the tip from being overtightened 2 Examine that the tip is not worn out Use a tip check ...

Страница 37: ...ngs on the connector are in good condition Replace O rings if worn O rings can be lubricated with a glycerine based lubricant 2 If you have an additional polisher tubing try replacing the polisher tubing 3 If the problem still remains contact your dealer for support and indicate Error Code E X02 8 11 Not sufficient or no water flow Polisher 1 Make sure that the Air Blow mode is not activated i e t...

Страница 38: ...ece 4 If the problem stays contact your dealer for support 8 13 Water drips from the bottom of the device onto the table 1 The LM ProPower device has a built in water trap that removes humidity from the pressurized air coming from the compressor in the compressor room of your clinic The water trap condenses the humidity in the pressurized air into water which is then released below the device Usua...

Страница 39: ...m 230 Vac Power cable Separate with protective earth plug Power consumption Max 40 VA Scaler power consumption Max 24 VA Scaler power output Max 10 W 24 kHz 28 kHz automatic tuning Ambient temperature Transport and storage 40 C to 70 C 40 F to 158 F Operation 10 C to 40 C 50 F to 104 F Relative humidity Transport and storage 10 to 100 Operation 10 to 95 Water supply pressure version conn to tap wa...

Страница 40: ...factory by the Authorized Dealer and Manufacturer s inspection establishes the existence of such a defect The customer must contact the Authorized Dealer from whom the products were purchased to request repair or replacement under this warranty and a written claim of the defects and send the original Product the Authorized Dealer This warranty is void if service or repair is performed by persons n...

Страница 41: ...LED Water connection Bottle Tap water connection Device Serial Number Purchase order number Order date Delivery date Installation Information Installation done date Installation done by person Installation location address Installation location building and room number ...

Страница 42: ...er and bottle in place according to instructions in section 3 8 Medicament bottle optional on page 14 Water supply tap water model Tap water pressure is ____ and according to the specification 1 10 bar 0 1 1 0 MPa Water connection is close knit and no indications of leakage is seen or heard Foot control Cable connected to the main unit Powder container Filled with LM ProPower powder and connected ...

Страница 43: ...aling with water mode Select WM3 100 and test foot pedal response Use artificial denture in this test Test dry mode Select 40 mode and test the foot pedal response Use artificial denture in this test Test dry mode Select 70 mode and test the foot pedal response Use artificial denture in this test Test dry mode Select 100 mode and test the foot pedal response Use artificial denture in this test Tes...

Страница 44: ...uments Oy PL 88 Norrbyn rantatie 8 FI 21601 Parainen Finland Tel 358 2 4546 400 Fax 358 2 4546 444 info lm dental com www lm dental com User manual language versions available at materialbank lm dental com 0915_9770090K ...

Отзывы: