Technical data
39
9. TECHNICAL DATA
Manufacturer
LM-Instruments Oy, PL 88 (Norrbyn rantatie 8) FI-21601 Parainen,
Finland
Model
LM-ProPower CombiLED
Classification
EN60601-1: Class 1, Type B
93/42/EEC: Medical products, class IIa
L x W x H
270 x 110 x 165 mm (without bottle and powder container)
Weight
3400 g
Voltage
100 Vac, 50-60 Hz
115 Vac, 50-60 Hz
230 Vac, 50-60 Hz
Primary fuse
T500 mA, 250 V, Ø5x20 mm (100 Vac)
T400 mA, 250 V, Ø5x20 mm (115 Vac)
T200 mA, 250 V, Ø5x20 mm (230 Vac)
Power cable
Separate with protective earth plug
Power consumption
Max. 40 VA
Scaler power consumption
Max. 24 VA
Scaler power output
Max. 10 W (24 kHz - 28 kHz, automatic tuning)
Ambient temperature
Transport and storage -40°C to 70°C (-40°F to 158°F)
Operation 10°C to 40°C (50°F to 104°F)
Relative humidity
Transport and storage 10% to 100%
Operation 10% to 95%
Water supply pressure (version
conn. to tap water)
1 - 10 bar (0.1–1,0 MPa, 14.5–145 PSI)
Water consumption
10 - 50 ml/min
Bottle volume (bottle version)
500 ml
Air supply pressure
4 - 10 bar (0.4 - 1.0 MPa, 58 - 145 PSI) Use only dry and filtered
compressed air.
Air consumption
Max. 20 l/min
Powder container capacity
40 g