Cleaning and maintenance
33
End of the day
• Run the automatic cleaning cycle with clean water for both the scaler and polisher handpiece. See
instructions “7.2. Automatic cleaning function for the scaler and polisher” on page 30.
• Remove and clean medicament bottle, bottle cap and powder container at a maximum temperature of
65 °C.
• Clean of the powder container cap with a soft cloth. Use a surface disinfectant suitable for hard plastics.
• Empty the powder container (4) and if some stays left, clean the O-ring (1) and cap with dry
compressed/pressurized air.
1
2
4
3
Weekly (bottle version)
• Run the automatic cleaning cycle with an anti-microbial cleaning agent solution in the bottle, for both
the scaler and polisher. See instructions “7.2. Automatic cleaning function for the scaler and polisher”
on page 30. We recommend to use a separate bottle for the cleaning agent solution. Concerning
exposure times of cleaning agent, follow instructions given by manufacturer.
• Before patient treatment. To flush the lines from cleaning agent solution, put clean water in the bottle and
run the automatic cleaning cycle for both the scaler and polisher. Do automatic cleaning cycle again until
clean water comes out of the handpiece.