Troubleshooting CombiLED
36
8.4. Weak or no LED light
1. Examine that lens in the LM-ErgoGrip is clear and clean. Clean or replace if necessary.
2. If you have an additional scaler handpiece (or polisher tubing), try replacing the scaler handpiece (or
polisher tubing).
3. If the problem stays – contact your dealer for support and indicate Error Code “E-S03 No light in scaler
handpiece“ or “E-P03 No light in polisher handpiece”.
8.5. No tip vibration
1. Make sure that the tip is tightly tightened. Preferably use a torque wrench and tighten clockwise until the
torque wrench slides and make sure that the tip is fully tightened. (The torque wrench secures a correct
tightening of the tip.)
2. Examine that the green light below the keys on the control panel is lit. If it is not lit, please see Section
8.1.
3. Make sure that the device has been set on by pressing the S-key and that the blue light adjacent to the
S-key is lit.
4. Make sure that the Working Mode has been selected by pressing one of the Working Mode keys and
that one of the yellow lights on the control panel is lit.
5. If you have an additional handpiece, try replacing the handpiece.
6. If you have an additional foot control cable, try replacing the foot control cable.
7. If you have an additional foot control, try replacing the foot control.
8. If the problem stays – contact your dealer for support and indicate Error Code “E S01 No vibration” or
“E-S02 Low power / vibration”.
8.6. Weak tip vibration
1. Make sure that the tip is tightly tightened. We recommend that you use a torque wrench and tighten
clockwise until the torque wrench slides. (The torque wrench prevents the tip from being overtightened.)
2. Examine that the tip is not worn out. Use a tip check card to evaluate wearing or try an unused tip. Use
original LM instruments tips for reliable performance.
3. If you have an additional handpiece, try replacing the handpiece.
4. If the problem stays – contact your dealer for support and indicate Error Code E-S02.
8.7. Not sufficient or no water flow (Scaler)
1. Make sure that the Dry Mode is not activated i.e. that the blue light adjacent to the crossed-out water
drop is not lit. If it is lit, press the S-key.
2. Adjust the water control ring on the handpiece to fully open (dots aligned).
3. If the unit are equipped with a water bottle: Make sure that the medicament/water bottle is correctly
connected i.e. the cap is tightened and pushed in all the way onto the connector. Examine that the
O-rings on cap and connector are in good condition. Replace O-rings if worn. O-rings can be lubricated
with a water soluble lubricant.
4. If the unit is connected with tap water: Make sure that the tap water hose is correctly connected to the
back of the device and that the tap water system is OK (tap/valve is open and possible filter is OK).
5. Try with different tip.