![LM-Dental LM-ProPower CombiLED Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/lm-dental/lm-propower-combiled/lm-propower-combiled_user-manual_1932118030.webp)
Cleaning and maintenance
30
7. CLEANING AND MAINTENANCE
NOTICE!
Immediately after using any type of medicament in the medicament bottle, run
the automatic cleaning cycle with clean water in the medicament bottle for both
the scaler and the polisher until clean water comes out of the handpieces.
Before cleaning and sterilizing, open water flow fully
open of the scaler handpiece and the polisher
(see images).
7.1. Air purge quick cleaning function
With the polisher mode active, a press on the C key will air purge the handpiece for some seconds. Air purge
the handpiece after each treatment to prevent clogging.
7.2. Automatic cleaning function for the scaler and polisher
NOTICE!
The automatic cleaning function must always be performed with the polisher
nozzle installed in the handpiece.
1. Set OFF the scaler or polisher by pressing the “S” or “P” key for 3 seconds. The blue function indicator
sets off.
2. Open the water control completely.
3. Put the scaler and/or the polisher handpiece above the cuspidor bowl.
4. Press first the cleaning key “C” and then the “S” and/or “P” key.
5. The cleaning cycle starts and stops automatically after 80 seconds.