Lighthouse NanoCount 30+ Operating Manual
A-12
248083427-1 Rev 4
Valid Data in Channels (30073 - 30076)
Register 30073 represents the flag bits corresponding to valid data
present in the analog register range. The mapping is such that bit 0 set
to TRUE (=1) would correspond to valid data present in Analog
Channel 1.
Register 30074 represents the flag bits corresponding to valid data
present in the particle register range.
Register 30075 represents the flag bits corresponding to analog
channels that have exceeded the threshold [Threshold High Registers
(45xxx series)] based on alarm mode.
Register 30076 represents the flag bits corresponding to particle
channels that have exceeded the threshold [Threshold High Registers
(45xxx series)] based on alarm mode
Data Type Registers
Note:
All data records
have the same data types
assigned to them. The
user does not have to read
the data type registers for
every record.
The 41xxx register series is used to identify the type of data items in the
30xxx series. The Data Type registers run in parallel with the Data
Registers. For example, Data Register 30041’s Data Type register is
41041.
Data Types are assigned 4 ASCII characters across 2 registers. If a Data
Type string contains less than 4 characters, then the rest of the string is
padded with NULL characters. Note that a Data Type using all four
characters will not end with a NULL character.
Содержание NanoCount 30+
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions NanoCount 30 Liquid Particle Counter Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse NanoCount 30 Operating Manual ii 248083427 1 Rev 4...
Страница 70: ...Lighthouse NanoCount 30 Operating Manual 6 6 248083427 1 Rev 4...
Страница 86: ...Lighthouse NanoCount 30 Operating Manual B 2 248083427 1 Rev 4...
Страница 93: ...00...
Страница 94: ......