Communications
248083427-1 Rev 4
4-21
15. Type
N
in response to “Change telnet config password”. The
instrument is shipped without a password. Changing the telnet
password or providing one is discouraged, except for absolute
security requirements. If a password is applied and forgotten, the
instrument must be returned to Lighthouse to clear the password.
See Figure 4-22.
Note:
When the
Ethernet restarts, it will
lose connection to the PC,
reported as, “Connection
to host lost”.
Figure 4-22 Telnet Config Password Screen
16. Type
S
to Save and restart and press
Enter
to save the changes.
See Figure 4-23.
Figure 4-23 Saving the Settings
17. Perform Step 7 though Step 16 for each
NC30+
.
WARNING:
Step 18 is
very important to prevent
network address errors
on the PC when the
process is complete.
18. Clear the ARP Cache by type
arp –d *
and press
Enter
.
If this is
not done before quitting this process, the PC may create “ghost
IPs” on the LAN and cause serious problems with the network.
19. Type
Exit
to quit the command console.