Communications
248083427-1 Rev 4
4-5
The
NC30+
complies with EIA’s RS-485 standards as shown in Table
4-1for distances and number of devices.
Refer to “Programming with MODBUS Protocol” on page 6-1 for
specific information on using MODBUS to program and communicate
with the instrument.
Table 4-1 EIA Industry Standards for RS-485 Communications
SPECIFICATIONS
RS-485
Mode of Operation
Differential
Total Number of Drivers and Receivers on
One Line (One driver active at a time for
RS85 networks)
32 Drivers
32 Receivers
Maximum Cable Length
4000 ft. (1,219.2 m)
Maximum Data Rate (40 ft. - 4000 ft. for
RS422/RS-485)
10Mb/s - 100Kb/s
Maximum Driver Output Voltage
-7V to +12V
Driver Output Signal Level (Loaded
Min.): LOADED
+/-1.5V
Driver Output Signal Level (Loaded
Max.): UNLOADED
+/-6V
Driver Load Impedance (Ohms)
54
Max Driver Current in High Z State
(POWER ON)
+/-100
μ
A
Max Driver Current in High Z State
(POWER OFF)
+/-100
μ
A
Receiver Input Voltage Range
-7V to +12V
Receiver Input Sensitivity
+/-200mV
Receiver Input Resistance (Ohms), (1
Standard Load for RS-485)
>12k
Содержание NanoCount 30+
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions NanoCount 30 Liquid Particle Counter Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse NanoCount 30 Operating Manual ii 248083427 1 Rev 4...
Страница 70: ...Lighthouse NanoCount 30 Operating Manual 6 6 248083427 1 Rev 4...
Страница 86: ...Lighthouse NanoCount 30 Operating Manual B 2 248083427 1 Rev 4...
Страница 93: ...00...
Страница 94: ......