Lighthouse NanoCount 30+ Operating Manual
A-8
248083427-1 Rev 4
Data
Registers
Data Registers
The Data input registers (43xxx series) are read/write. All enable data
items are 4 bytes long and are stored across 2 registers. Byte and word
ordering is big-endian. Thus, data items are formed by placing the high
bytes in front of the low bytes. For example:
<High Bytes><Low Bytes> = <4 Byte Data Item>
The 43xxx register series is used to determine which particle data
channel is ENABLED. These registers supersede the older Data Enable
Registers (31xxx) which have been obsoleted.
These registers run in parallel with the data registers (30xxx series).
For example, data register 30010’s enable register would be 43010.
Data register 30016’s enable register would be 43016.
Table A-6 Enable/Disable Bits
Bit
Description
0
DATA ENABLE (0=disable; 1=enable)