Lighthouse NanoCount 30+ Operating Manual
4-4
248083427-1 Rev 4
Connecting to a PC
Refer to Figure 4-4 illustrating the LWS RS-485 Converter Kit.
Figure 4-4 LWS RS-485 Converter Kit Example
1. Disconnect power from the instrument.
2. Connect one end of a standard Cat5 cable to an LWS RS-485
converter.
3. Connect the other end of the Cat5 cable to the Communication port
on the instrument.
4. Connect the RS-232, DB9 side of the RS-485 converter to a DB9
Male to DB9 Female straight through cable. Connect the female
DB9 to any available COM port on the computer.
5. Power on the instrument.
Connecting to an RS-485 Network
To connect a PC to multiple units or RS-485 network, use a Cat5 cable
and RS-485 adapter to attach the computer to an RS-485 hub. Connect
the instrument to the hub using the Industrial RJ45 cable referred to
starting in “Cable Build” on page 3-2.
Make sure that each
NC30+
or other RS-485 instrument has a unique
address (non-zero) or a conflict will result and data will be lost.
RS-485 Converter &
Power Supply
RS-232 Male-to-Female
Cable