Getting Started
248083427-1 Rev 4
3-7
11. Insert the RJ45 connector into a crimp tool similar to that shown in
Figure 3-6. Maintain inward pressure while crimping the wires into
their channels.
Figure 3-6 Typical RJ45 Crimp Tool
12. Check the cable wire ends in the RJ45 to make sure they have been
crimped properly. Removing the connector at this point is a lot
easier than after everything is installed and communication with the
instrument is failing due to a poor crimp. Review Figure 3-7 for an
illustration of how tightly the wire ends should be against the ends
of each channel.
Figure 3-7 Close-up of RJ45 Wire Detail
Wires against connector end
Note that wires are
not against end
Содержание NanoCount 30+
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions NanoCount 30 Liquid Particle Counter Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse NanoCount 30 Operating Manual ii 248083427 1 Rev 4...
Страница 70: ...Lighthouse NanoCount 30 Operating Manual 6 6 248083427 1 Rev 4...
Страница 86: ...Lighthouse NanoCount 30 Operating Manual B 2 248083427 1 Rev 4...
Страница 93: ...00...
Страница 94: ......