Rev. D
Performance Verification
5-45
5.11.2 Trigger on Pulse Width 100 nsec
a.
Pulse Width < 100 nsec
!
Recall
xxxP047.PNL
or configure the DSO as shown in 5.11.1.a
and make the following changes :
Width < 100 nsec :
On
Width > 100 nsec
:
Off
Time base
:
20 nsec/div.
!
Set the generator frequency to
10 MHz
.
!
Check that the scope Triggers. Record the test result in Table 12.
!
Set Width < 100 nsec
Off
and
Width > 100 nsec
ON
!
Check that the scope
doesn’t trigger :
slow trigger and no flashes in box next
to normal. Record the test result in Table 12.
b. Pulse Width > 100 nsec
!
Set the generator frequency to
4 MHz
.
!
Time base
:
50 nsec/div.
!
Set Width < 100 nsec
Off
and
Width > 100 nsec
ON
!
Check that the scope Triggers. Record the test result in Table 12.
!
Set Width < 100 nsec
On
and
Width > 100 nsec
Off
!
Check that the scope
doesn’t trigger :
slow trigger and no flashes in box.
Record the test result in Table 12.
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...