Performance Verification
5-25
b. DC 50
Ω
Ω
Ω
Ω
with Bandwidth Limiter On
!
Recall
xxxP024.PNL
or configure the DSO:
Panel Setups
:
Recall FROM DEFAULT SETUP
Channels Trace ON
Channel 1
Input Coupling
:
DC 50
Ω
Ω
Ω
Ω
Global BWL
:
20 MHz
Input gain
:
100 mV/div.
Input offset
:
0 mV
Trigger setup
:
Edge
Trigger on
:
1
Slope line
:
Pos
Mode
:
Auto
Time base
:
1
µµµµ
sec/div
.
Channel use
:
4
Record up to
:
25 k
WAVE PILOT
:
MEASURE
Mode
:
Custom
Statistics
:
Off
Change parameters
On line 1
:
Sdev of 1
On line 2
:
Freq of 1
!
Connect the test equipment as shown in Figure 5-6.
!
Set the generator frequency to
300 kHz
.
!
Adjust the generator signal amplitude to measure
sdev(1)
=
200 mV.
!
Set Time base
:
50 nsec/div
.
!
Increase the generator frequency until
sdev(1)
=
140 mV.
(typically 25 MHz)
!
Press
Clear Sweeps
!
When
sdev(1)
=
140 mV,
record Freq(1) in Table 9.
!
Check that the frequency is within the limits specified in Table 9.
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...