Rev. D
Performance Verification
5-21
!
Connect a 5ns 50
Ω
BNC cable to the
RF output
of the HP8648B
generator and then through a 6dB attenuator and the necessary adapters to the
power sensor.
Figure 5-5 : Power Meter Equipment Setup
!
Set the generator frequency to
300 kHz
!
Set the power meter to the correct
Correction Factor
!
Set the generator amplitude to measure
0.200 mW
on the power meter.
!
Read the displayed
generator output amplitude
, and record it in the third
column of Table 7.
!
Repeat the above measurement for 1.1 MHz, 10.1 MHz, 100.1 MHz, 250.1 MHz
& 500.1 MHz (changing the correction factor of the power meter at each
frequency). Record the generator output amplitude readout in the third column of
Table 7 for LT37X or LT354. For LT26X, use Table 7A for frequencies of 1.1
MHz, 10.1 MHz, 100.1 MHz, 175.1 MHz & 350.1 MHz. For LT584, use Table 7B
for frequencies of 1.1 MHz, 10.1 MHz, 100.1 MHz, 250.1 MHz, 500.1 MHz &
1000.1 MHz.
!
Disconnect the
RF output
of the HP8648B generator from the
HP8482A power
sensor.
!
Connect
the
RF output
of the HP8648B generator through a
5ns 50 Ohm BNC cable and a 6 dB attenuator into Channel 1.
!
Set the generator frequency to
300 kHz
.
!
From the generator, apply the
recorded
generator signal amplitude
to
Channel 1.
!
Press
Clear Sweeps
.
Power Sensor
Power Meter
Power Ref
Output
Sensor Input
Sine Wave
Generator
6
DB
6 db
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...