Rev. D
Performance Verification
5-31
5.10.2 Channel Trigger at +3 Divisions Threshold
DC Coupling
!
Recall
xxxP032.PNL
or configure the DSO as shown in 5.10.1.a and for
each Channel make the following change :
Set Trigger level :
DC +300 mV
!
Connect the output of the generator to Channel 1 through a 50 Ohm coaxial
cable.
!
Press
Clear Sweeps
,
!
Acquire 10 sweeps and record in Table 10 the
level
readout displayed below
100 mV in the icon
1,
at top left
.
!
Compare the test results to the corresponding limit in the test record.
!
Set Trigger Slope 1 :
Neg
!
Acquire 10 sweeps and record in Table 10 the
level
readout displayed below
100 mV in the icon
1,
at top left
.