Rev. D
Performance Verification
5-27
!
Increase the generator frequency until
sdev(1)
=
140 mV.
(typically 200 MHz)
!
Press
Clear Sweeps
!
When
sdev(1)
=
140 mV,
record Freq(1) in Table 9.
!
Repeat the 20 MHz and 200 MHz Bandwidth limiter tests for the other channels,
substituting channel controls and input connector.
!
Recall
xxxP025.PNL
for Channel 2
, xxxP026.PNL
for Channel3
xxxP027.PNL
for Channel 4, or configure the DSO as shown in 5.9.1
Procedure and make the necessary changes.
!
Record the test results in Table 9, and compare the results to the limits.