Rev. D
Performance Verification
5-5
!
For all input channels measure the
input impedance
.
!
Record the input impedance in Table 2, and compare it to the limits.
!
Recall
xxxP004.PNL
or Set Input gain to
200 mV/div
on all 4 Channels.
!
Repeat the test for all input channels.
!
Record the measurements in Table 2, and compare the results to the limits in
the test record.
c.
DC 50
Ω
Ω
Ω
Ω
!
Recall
xxxP005.PNL
or configure the DSO as shown in 5.4.1.a,
and for each Channel make the following change:
Input Coupling
:
DC 50
Ω
Ω
Ω
Ω
!
For all input Channels, measure the input impedance.
!
Record the
input impedance
in Table 2, and compare it to the limits.
!
Recall
xxxP006.PNL
or set Input gain to
200 mV/div.
on all 4 Channels
!
Repeat the test for all input channels. Record the measurements in Table 2,
and compare the results to the limits in the test record.
5.4.2
External Trigger Input Impedance
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...