5-38
Performance Verification
Rev. D
DC Coupling
!
Recall
xxxP042.PNL
or configure the DSO as shown in 5.10.4.a and make
the following change :
Set Ext Trigger level
:
DC
−−−−
300 mV
!
Connect the test equipment as shown in Figure 5-7.
!
Press
Clear Sweeps
.
!
Acquire 10 sweeps and record in Table 11 the
level
readout displayed below
100 mV in the icon
2,
at top left
.
!
Compare the test results to the corresponding limit in the test record.
!
Set Trigger Slope Ext
:
Neg
!
Acquire 10 sweeps and record in Table 11 the
level
readout displayed below
100 mV in the icon
2,
at top left
.
5.10.7 External/10 Trigger at 0 Division Threshold
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...