Rev. D
Performance Verification
5-3
5.4
Input Impedance
The impedance values for 50
Ω
and 1M
Ω
couplings are measured with a high
precision digital multimeter. The DMM is connected to the DSO in 4-wire
configuration (input and sense), allowing for accurate measurements.
Specifications & Test limits
DC 1.00 M
Ω
±
1 %
AC 1.20 M
Ω
±
1 % (2mV/div to 99mV/div)
1.00 M
Ω
±
1 % (100mV/div to 10V/div)
DC 50
Ω
±
1 %
5.4.1
Channel Input Impedance
a. DC 1M
Ω
Ω
Ω
Ω
!
Recall
xxxP001.PNL
or configure the DSO:
Panel Setups
:
Recall FROM DEFAULT SETUP
Channels Trace OFF
Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling
:
DC 1M
Ω
Ω
Ω
Ω
on all 4 Channels
Input gain
:
50 mV/div.
on all 4 Channels
Time base
:
50 nsec/div.
Trigger mode
:
Auto
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...