5-20
Performance Verification
Rev. D
5.9
Bandwidth
The purpose of this test is to ensure that the entire system meets the bandwidth
specification. An external source is used as the reference to provide a signal
where amplitude and frequency are well controlled.
If a leveled sine wave generator is not used then the amplitude of the generator as
a function of frequency and power must be calibrated using an HP8482A sensor on
an HP437B power meter or equivalent.
Specifications & Test limits
DC to 1GHz (10mV to 5V/div)
LT584
DC to 500MHz (10mV to 5V/div)
LT37x, LT354
DC to 350MHz (10mV to 5V/div)
LT26x
Procedure
!
Recall
xxxP022.PNL
or configure the DSO
Panel Setups
:
Recall FROM DEFAULT SETUP
Channels Trace ON
Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling
:
DC 50
Ω
Ω
Ω
Ω
on all 4 Channels
Input gain
:
50 mV/div
on all 4 Channels
Input offset
:
0 mV
on all 4 Channels
Trigger setup
:
Edge
Trigger on
:
Line
Slope line
:
Positive
Mode
:
Auto
Time base
:
1
µµµµ
sec/div
.
Channel use
:
4
Record up to
:
25 k
WAVE PILOT
:
MEASURE
Mode
:
Custom
Statistics
:
On
Change parameters
On line 1
:
Sdev of 1
On line 2
:
Sdev of 2
On line 3
:
Sdev of 3
On line 4
:
Sdev of 4
!
Connect the HP8482A power sensor to the power meter.
!
Zero and
calibrate
the power sensor according to the power meter instructions.
!
Connect a
BNC adapter
to the power sensor.
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...