Rev. D
Performance Verification
5-15
On line 2
:
Measure mean of B
On line 3
:
Measure
mean of C
On line 4
:
Measure mean of D
!
For the low sensitivities:
2 mV, 5 mV, 10 mV and 20 mV/div
., connect the test
equipment as shown in Figure 5-1.
Figure 5-1: DC 1M
Ω
Ω
Ω
Ω
Accuracy Equipment Setup for 2,5,10 and 20 mV/div.
!
For 50 mv/div & 100 mV/div, connect the test equipment as shown in Figure 5-2.
!
For 5V/div, no attenuator is required, connect the test equipment as shown in
Figure 5-3.
Figure 5-2: DC 1M
Ω
Ω
Ω
Ω
Accuracy Equipment Setup for 100 mV/div
Figure 5-3 : DC 1M
Ω
Ω
Ω
Ω
Accuracy Equipment Setup for 5V/div.
!
For each
DSO Volts/div
, set the output of the external
DC voltage reference
source
as shown in Table 5, column PS output.
DC Power Supply
DMM
DC Power Supply
DMM
20
DB
20
DB
20db 20db
DC Power Supply
DMM
20
DB
20db
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...