Rev. D
Performance Verification
5-9
5.5.2
External Trigger Leakage Current
a. DC 50
Ω
Ω
Ω
Ω
!
Recall
xxxP011.PNL
or configure the DSO as shown in 5.5.1
and make the following changes :
Select Setup trigger
Set Trigger on
:
EXT
External
:
DC 50
Ω
Ω
Ω
Ω
!
Connect the DMM to External.
!
Measure the
voltage
and enter it in Table 3. Compare it to the limits.
Содержание Waverunner2 LT262 Series
Страница 1: ...LTXXX2 SM E LeCroy Waverunner2 Series Service Manual Version D December 2003 ...
Страница 30: ...4 12 Theory of Operation Block diagram 1 HAM631 ...
Страница 55: ...Rev D Performance Verification 5 19 ...
Страница 62: ...5 26 Performance Verification Rev D Set Global BWL 200 MHz Set Timebase 5 nsec div ...
Страница 80: ...5 44 Performance Verification Rev D a Pulse Width 10 nsec b Pulse Width 10 nsec ...
Страница 84: ...5 48 Performance Verification Rev D ...
Страница 86: ...Rev D 2 of 12 This page intentionally left blank ...
Страница 96: ...Rev D 12 of 12 LT584 LT374 372 LT354 LT264 262 Test Record This page intentionally left blank ...
Страница 114: ...6 18 Maintenance ...
Страница 126: ...7 12 Mechanical Parts Removal F2 ...
Страница 154: ...Mechanical Part Removal 7 21 ...