SoundAdvisor Model 831C
Setting Up Cloud Storage for Data Files
9-14
FIGURE 9-9
Alert Notification Email Example
Spam and Duplicate Alerts
To protect recipients from receiving duplicate alert notifications for the
same trigger when the state of the meter fluctuates around that trigger,
the 831C has a hysteresis, or threshold, for multiple notifications that
concern these specific alert details:
• Memory: If a notification is sent, free memory space can raise up
to 5% of the trigger level before another alert can be sent.
• Temperature: If a notification is sent, the temperature can rise
above the trigger (-40°C) or fall below the trigger (70°C) multiple
times until the temperature rises/falls at least 5°C. If the same
conditions are met again, after that, another alert notification is
sent.
• Low external power: After an alert is sent, the voltage can rise
up to 0.5 V before another alert can be sent if the conditions are
met.
9.12 Setting Up Cloud Storage for Data Files
When a measurement data file is automatically or manually saved, you
can save it on the meter, to a USB drive, or upload directly to a remote
storage service such as Dropbox
1
or an SFTP account. You can also
1. Dropbox is a registered trademark of Dropbox, Inc. in the US and other
countries.