![Larson Davis HVM100 Скачать руководство пользователя страница 142](http://html1.mh-extra.com/html/larson-davis/hvm100/hvm100_manual_3699279142.webp)
C-10
HVM100 User Manual
4/11/03
Read Time History Commands
The read time history commands are used to read data from
the time history buffer. The syntax for this command is
Hcnnn,i. The c indicates which channel to read (X, Y, Z, or
S). The nnn is an index for indicating which sample to read.
Example:
nnn = 0 is the last sample stored, nnn = 1, is the next to last sample
stored.
The i indicates which data to read (i = 0 for Arms, or i = 1
for Peak). The commands listed in the following table are
also available. The date information (H0 - H5) refers to the
starting date/time of the first history record.
Hint:
The HVM100 always returns data in decibels, referenced to (10
-6
m/s
2
).
Command
Parameter
Settings
H0
History: Year
YY (00 to 99, 99 = 1999, 00 = 2000, 98 =
2098)
H1
History: Month
mm (01 to 12)
H2
History: Day
dd (01 to 31)
H3
History: Hour
hh (00 to 23)
H4
History: Minute
mm (00 to 59)
H5
History: Second
ss (00 to 59)
H6
Number of Samples
000 to 239
H7
Number of Wraps
Returns the number of times the buffer has
been completely filled. After the buffer is
filled, new data is simply written over the
existing data. (i.e. the buffer always contains
the last 240 samples, or 120 samples if Peak is
also stored.)
Содержание HVM100
Страница 8: ...HVM100 User Manual 6...
Страница 18: ...1 10 HVM100 User Manual 4 11 03...
Страница 29: ...4 11 03 Getting Started 2 11 Weighting X Wk...
Страница 30: ...2 12 HVM100 User Manual 4 11 03...
Страница 44: ...3 14 HVM100 User Manual 4 11 03...
Страница 50: ...4 6 HVM100 User Manual 4 11 03...
Страница 55: ...4 11 03 Printing 5 5 Step 5 Select Hyperterminal from the menu A new con nection dialog box will appear...
Страница 56: ...5 6 HVM100 User Manual 4 11 03 Step 6 Enter a name and choose an icon for your Hyper terminal connection...
Страница 57: ...4 11 03 Printing 5 7 Step 7 Press the OK button...
Страница 58: ...5 8 HVM100 User Manual 4 11 03 Step 8 The Connect to dialog box will appear...
Страница 62: ...5 12 HVM100 User Manual 4 11 03 Left click the Properties button Properties Settings...
Страница 63: ...4 11 03 Printing 5 13 Left click the Settings Tab at the top of the window to open the Setting dialog box...
Страница 67: ...4 11 03 Printing 5 17...
Страница 68: ...5 18 HVM100 User Manual 4 11 03...
Страница 80: ...6 12 HVM100 User Manual 4 11 03...
Страница 93: ...4 11 03 Powering the HVM100 8 5 The connector pinout is as follows Negative Positive Positive Negative...
Страница 94: ...8 6 HVM100 User Manual 4 11 03...
Страница 102: ...9 8 HVM100 User Manual 4 11 03...
Страница 146: ...D 2 HVM100 User Manual 4 11 03...
Страница 147: ...4 11 03 D 3...
Страница 148: ...D 4 HVM100 User Manual 4 11 03...
Страница 149: ...4 11 03 D 5...
Страница 150: ...D 6 HVM100 User Manual 4 11 03...
Страница 151: ...4 11 03 D 7...
Страница 152: ...D 8 HVM100 User Manual 4 11 03...
Страница 153: ...4 11 03 D 9...
Страница 154: ...D 10 HVM100 User Manual 4 11 03...
Страница 155: ...4 11 03 D 11...
Страница 156: ...D 12 HVM100 User Manual 4 11 03...
Страница 157: ...4 11 03 D 13...
Страница 158: ...D 14 HVM100 User Manual 4 11 03...
Страница 170: ...F 4 Warranty Customer Satisfaction April 11 2003...
Страница 174: ...HVM100 User Manual 4...