4/11/03

HVM100

Human Vibration Meter

Larson Davis

1681 West 820 North, Provo, UT 84601-1341

Phone: (801) 375-0177, FAX: (801) 375-0182

www.larsondavis.com

IHVM100.01 Rev. D

Содержание HVM100

Страница 1: ...4 11 03 HVM100 Human Vibration Meter Larson Davis 1681 West 820 North Provo UT 84601 1341 Phone 801 375 0177 FAX 801 375 0182 www larsondavis com IHVM100 01 Rev D...

Страница 2: ...performance merchantability or fitness for a particular purpose This documentation is subject to change without notice and should not be construed as a commitment or representation by Larson Davis Su...

Страница 3: ...1 Menu Navigation 2 1 Alpha Numeric Menu Function 2 1 Indexed Menu Function 2 3 Key Board Functions 2 4 Setup Key 2 5 Mode Selection 2 5 Averaging Time 2 6 Time History Settings 2 6 Store Time 2 7 Au...

Страница 4: ...terminal 5 3 Chapter 6 Storing and Recalling Data Files and Setups 6 1 STORE Key 6 1 Storing a Setup 6 1 Storing a File 6 4 Temporary Data Storage File Register 00 6 5 Data Storage File Registers 01 t...

Страница 5: ...dapter 9 5 AC DC Output Connector 9 6 Pinout 9 7 Appendix A Specifications A 1 General Characteristics A 1 Type Precision A 1 Effects of Temperature A 2 Effects of Humidity A 2 Limits of Temperature a...

Страница 6: ...16 A 17 WB Whole Body Frequency Weighting A 17 Typical Measurements Ranges A 19 Vibration Fa A 19 Vibration Fb Fc Ws A 20 Hand Arm Wh A 21 Whole Body WB A 22 Whole Body Wc A 23 Whole Body Wd We A 24...

Страница 7: ...s C 3 Read Data Commands C 9 Read Time History Commands C 10 Control Commands C 11 Appendix D Frequency Response Curves D 1 Appendix E Miscellaneous Information E 1 DC Output Calibration E 1 AC DC Out...

Страница 8: ...HVM100 User Manual 6...

Страница 9: ...cy weighting and band limiting settings single and double integration dis played data in a variety of units and independent AC or DC outputs for all 3 channels simultaneously We invite you to read thi...

Страница 10: ...from the HVM100 Chapter 6 Storing and Recalling Data Files and Setups This chapter will describe in detail how to store data and setup information in the HVM100 s on board memory Chapter 7 Taking Mea...

Страница 11: ...transducer is especially critical to obtaining good results Standard Accessories The HVM100 is delivered with the following standard accessories User Manual Batteries 2 x AA IEC Type LR6 Optional Acc...

Страница 12: ...pin Microtech to 4 pin Lemo female CBL123 Cable 6 pin DIN male to 4 pin Lemo female CBL124 Cable AC DC Output 3 X BNC to Switchcraft CBL125 Cable Mini 4 pin to 4 pin Lemo female ADP060 Adapter hand a...

Страница 13: ...ld or factory environments with little signal degradation NOTE The HVM100 has built in ICP power supplies and charge amplifiers This will allow the HVM100 to interface directly with ICP or charge tran...

Страница 14: ...above the upper limit of the HVM100 and will subsequently be masked out by the low pass filter on the HVM100 If the res onant frequency becomes an issue it is recommended that an external mechanical f...

Страница 15: ...kHz g rms 0 005 0 0005 0 0001 Shock Limit g pk 7 000 7 000 7 000 Operating Temperature Range o F o C 65 to 250 54 to 121 65 to 250 54 to 121 65 to 176 54 to 80 Amplitude Linearity 1 1 1 Transverse Se...

Страница 16: ...ency Weighting Selections NOTE The Hand Arm and Whole body modes are separately purchased options These options and the frequency weight ings associated with these options will not be available unless...

Страница 17: ...ed with each of the frequency weightings Each band limiting filter is a combination of a high pass and low pass 2nd order Butterworth filter 12 dB per octave attenuation with 3 dB corner frequencies a...

Страница 18: ...1 10 HVM100 User Manual 4 11 03...

Страница 19: ...Check key to enter the modify section of the menu You can now use the r Right and l Left arrow keys to scroll through the options for that item After the appropriate selection is in the win dow use t...

Страница 20: ...tion When you move to the next position the HVM100 will start you from where you left off i e if you entered an F in the first position when you move to the next position you will be starting from the...

Страница 21: ...to change Step 2 Press the c key to place the HVM100 in the mod ify mode The selection will begin to blink Step 3 Use the l and r arrow keys to scroll through the selections in the menu r r Step 4 Whe...

Страница 22: ...t of the HVM100 Key Board Functions The keyboard on the front of the HVM100 is the main inter face to the instrument The keyboard consists of 11 dedi cated function keys 4 arrow keys and the check key...

Страница 23: ...etup menu To access the Setup menu press the SETUP key and then use the u and d arrow keys to navigate through the menu items NOTE To view each selection first press the c key and then press the r or...

Страница 24: ...data is col lected and averaged before a new average is started The selections are Slow 1 sec 2 sec 5 sec 10 sec 20 sec 30 sec and 60 sec The Slow response is a 1 second exponential detector while the...

Страница 25: ...an be set from 1 minute to 99 hours Auto Store Hint This setting is in the Setup menu To access the Setup menu press the SETUP key and then use the u and d arrow keys to navigate through the menu item...

Страница 26: ...s are full The Auto Store Autostop feature will allow you to take a fixed length measurement The Autostop feature will stop the meter when the Store time is reached just as if you had pressed the s ke...

Страница 27: ...ow key to scroll through each selec tion In the vibration mode you can select from the following weighting selections Please see the appendix at the end of this manual for details on weighting curves...

Страница 28: ...OTE To view each selection first press the c key and then press the r or l arrow key to scroll through each selec tion In the Whole Body mode you can select from the following weighting selections as...

Страница 29: ...4 11 03 Getting Started 2 11 Weighting X Wk...

Страница 30: ...2 12 HVM100 User Manual 4 11 03...

Страница 31: ...nu To access the Tools menu press the TOOLS key and then use the u and d arrow keys to navigate through the menu items NOTE To view each selection first press the c key and then press the r or l arrow...

Страница 32: ...ground or have some other problem Check all connections cables and connectors This setting would be used to amplify the signal from a charge type accelerometer Display Units Hint This setting is in t...

Страница 33: ...nu items NOTE Integration is a mathematical process of changing the displayed units from acceleration to velocity to displace ment Single integration changes the dis played parameter from acceleration...

Страница 34: ...VM100 allows the user to select a different k factor for each channel The default setting is 1 00 however some applications may require different settings NOTE To change the Sum Factor parameter press...

Страница 35: ...tput is useful for a chart recorder or to measure the voltage pro portional to the RMS or peak value The DC output is updated according to the averaging time If the averaging time is set to slow the s...

Страница 36: ...el selections however the available selections will be the same for all operating modes as well The AC output will produce a signal from 0 0 5 Volt RMS The scale of the DC output is typically 5 mV dB...

Страница 37: ...M100 DC max This signal will be a DC voltage with a level proportional to the max level displayed on the HVM100 DC peak This signal will be a DC voltage with a level proportional to the peak level dis...

Страница 38: ...navigate through the menu items The Baud rate selection will allow you to communicate with the PC or the serial printer The choices are as follows NOTE To view each selection first press the c key and...

Страница 39: ...h this selection the HVM will only send the overall data from the 3 input channels and the sum channel Erase All Files Hint This setting is in the Tools menu To access the Tools menu press the TOOLS k...

Страница 40: ...he c key the screen will display the following message After about 3 seconds the display will return to the Erase All Files screen All file registers have now been cleared Erase Setups Hint This setti...

Страница 41: ...ressing the c key the screen displays the following message The display will return to the Erase Setups screen after about 3 to 10 seconds All Setups have now been cleared from the setup registers Hea...

Страница 42: ...cursor on the Left most position you want erased Press the R key and all positions to the right of the cursor including the cursor position will be erased If the cursor position is all the way to the...

Страница 43: ...en use the u and d arrow keys to navigate through the menu items NOTE To change the Date and Time parameter press the c key use the u and d arrows to select the desired value and then use the r and l...

Страница 44: ...3 14 HVM100 User Manual 4 11 03...

Страница 45: ...Auto Range function in the HVM100 is designed to work with a steady state signal It is not meant for use while taking data Before enabling this function the transducer should be attached to the stead...

Страница 46: ...hannel at a time and manually assign a gain set ting to that channel The choices are 0 20 40 or 60 dB NOTE To view each selection first press the c key and then press the r or l arrow key to scroll th...

Страница 47: ...lways be entered in g s rms During calibration the gain of the HVM100 should also be set so that the level of the calibrator is within the HVM100 s measurement range For example when using a 1 0 g rms...

Страница 48: ...Y and Calibrate Z displays are located just below the Calibrate X display From the Calibrate X display press the d arrow key once to get to the Calibrate Y display and press the d arrow key again to g...

Страница 49: ...ange The entry procedure is the same for all channels and the screens will look the same The units for ICP and Direct are mV g rms the units for charge accelerometers are pC g rms Reference Accelerati...

Страница 50: ...4 6 HVM100 User Manual 4 11 03...

Страница 51: ...o stream formatted ASCII text to the instrument s serial port connector NOTE The print key only works if you are currently in the History or Data menu Only the current data in the HVM100 can be printe...

Страница 52: ...a Printer Printing to a printer is the easiest way to download data from the HVM100 The printer must have a serial interface Some desktop laser printers will have a serial port on the back or serial p...

Страница 53: ...lled if a custom installation of Windows was performed This option will allow you to print the text into the Hyperterminal screen and then cut and paste text into a word processing or spreadsheet appl...

Страница 54: ...5 4 HVM100 User Manual 4 11 03 Step 4 Locate Hyperterminal on your computer Hyper terminal is usually found in the Start menu under Programs and in the Accessories folder...

Страница 55: ...4 11 03 Printing 5 5 Step 5 Select Hyperterminal from the menu A new con nection dialog box will appear...

Страница 56: ...5 6 HVM100 User Manual 4 11 03 Step 6 Enter a name and choose an icon for your Hyper terminal connection...

Страница 57: ...4 11 03 Printing 5 7 Step 7 Press the OK button...

Страница 58: ...5 8 HVM100 User Manual 4 11 03 Step 8 The Connect to dialog box will appear...

Страница 59: ...9 Step 9 The last selection in this box is the Connect using selection box This will allow you to select the Com port you will use to communicate with the HVM100 Select the Com port you plugged your H...

Страница 60: ...4 11 03 Step 10 The Communications properties screen will appear Select the Baud rate bits per second that matches the baud rate on your HVM100 The other settings are Data bits 8 Parity None Stop Bits...

Страница 61: ...the OK button You have now established a connection for the HVM100 You will notice that in the lower left corner of the Hyperterminal screen it states that there is a connection and will show how lon...

Страница 62: ...5 12 HVM100 User Manual 4 11 03 Left click the Properties button Properties Settings...

Страница 63: ...4 11 03 Printing 5 13 Left click the Settings Tab at the top of the window to open the Setting dialog box...

Страница 64: ...d click ASCI Setup to open the ASCI Idialog box Configure the items as shown above and press OK to close the ASCII dialog box Press OK to close the Properties window Left click Transfer box which will...

Страница 65: ...5 15 This will open the following menu Use the Browse function to identify the file into which you want the data transfered and press Start After the connection has been established press the PRINT k...

Страница 66: ...5 16 HVM100 User Manual 4 11 03 Left click Transfer highlight Stop on the drop down menu and left click The data can now be found as a Test file in the folder designated for saving the transfered file...

Страница 67: ...4 11 03 Printing 5 17...

Страница 68: ...5 18 HVM100 User Manual 4 11 03...

Страница 69: ...s It will allow you to store data files and setups The functionality is determined by the menu you are currently viewing If you press the store key while you are viewing the Data or His tory menu you...

Страница 70: ...register you are currently select ing NOTE To purge all setup registers go to the Tools menu use the u and d arrow keys to select the Erase Setups selection Press the c key use the r arrow key to cha...

Страница 71: ...u and d arrows to select characters for the positions and the r and l arrow keys to move to the different positions Step 6 Press the c key This will store the setup in the selected register and then...

Страница 72: ...a screen where data is being displayed Step 2 Press the STORE key Step 3 The Store File screen will appear The number in the bottom right of the screen is referring to the file register where the data...

Страница 73: ...ister The data file will also contain all of the setup information that corre sponds to that data file When the data file is recalled all of the setup information can also be viewed The data can now b...

Страница 74: ...ile register 02 and so on until all 99 file registers are full Preserving the File Counter At Power Off The file counter which keeps track of the last file register used is saved when the HVM100 is tu...

Страница 75: ...rent memory of the HVM100 The RECALL key is again dependent on the menu being viewed when the key is pressed If you are in the Tools Setup or Range menu and press the RECALL key you will be prompted t...

Страница 76: ...6 8 HVM100 User Manual 4 11 03 Step 1 Enter one of the three valid menus for recalling a setup The Range menu the Setup menu or the Tools menu...

Страница 77: ...ss the RECALL key Step 3 You will now be prompted to recall a setup Step 4 Use the u and d arrow keys to scroll through the setups until the desired setup appears on the screen Step 5 Press the c key...

Страница 78: ...d save the setup to regis ter S0 and the HVM100 will automatically recall that setup every time the HVM100 is turned on If for some reason the HVM100 needs to be reset to its orig inal factory default...

Страница 79: ...to recall a file Step 4 Use the u and d arrow keys to move to the desired selection Step 5 Press the c key to recall the file into the current memory All setup information was stored when the data fi...

Страница 80: ...6 12 HVM100 User Manual 4 11 03...

Страница 81: ...ent There are several screen indicators that will appear as this key is pressed Overload Detection When an overload occurs on the HVM100 the meter will perform three functions for the overloaded chann...

Страница 82: ...e bar graph run indicator whenever the channel currently displayed is under range The under range indicator like the bar graph is inde pendent for all three channels A channel is under range whenever...

Страница 83: ...of time history records is cut in half to 120 NOTE The 2nd History setting located in the Setup menu controls whether or not the Peak levels are stored in the Time History Example You have selected an...

Страница 84: ...7 4 HVM100 User Manual 4 11 03 The HISTORY key works much the same as the other menus in the HVM100 To view the Time History data press the HISTORY key...

Страница 85: ...data from the different channels u u u Arms 0 00 04 01900 m s2 FaZ Arms 0 00 03 01900 m s2 FaZ Arms 0 00 02 01900 m s2 FaZ Arms 0 00 01 01900 m s2 FaZ Arms 0 00 01 01900 m s2 FaZ This is the value be...

Страница 86: ...History turned on If the 2nd history selection is turned on the History buffer will alternate between RMS and Peak values for each record u u u Peak 0 00 02 4 3800 m s2 FaZ Arms 0 00 02 1 5000 m s2 F...

Страница 87: ...on first line RMS level units frequency weighting and channel on second line A acceleration Arms 10 sec 3 5700 m s2 FaZ 3rd screen reports minimum level on first line and max imum level on second line...

Страница 88: ...Crest factor CF is not calculated if the Averaging Time setting is SLOW The Frequency weighting and channel are also reported A 4 00052 A 8 00036 FaZ The 8th screen shows the Allowed Exposure Time ba...

Страница 89: ...Remaining Battery Voltage The battery voltage can be viewed from the data menu by pressing the c key The following screen will be displayed When the batteries have approximately five minutes of life r...

Страница 90: ...8 2 HVM100 User Manual 4 11 03 Step 1 Remove battery cover from the side of the case Step 2 Remove old batteries...

Страница 91: ...Step 3 Arrange new batteries according to diagram on the inside of the case Make sure polarity of the batter ies is correct Step 4 Gently push the new batteries in the case while pushing the battery...

Страница 92: ...olts DC external power The external power connector is on the top of the HVM100 The power supply normally used with the HVM100 is the Larson Davis PSA027 This is a switching power supply that will ope...

Страница 93: ...4 11 03 Powering the HVM100 8 5 The connector pinout is as follows Negative Positive Positive Negative...

Страница 94: ...8 6 HVM100 User Manual 4 11 03...

Страница 95: ...VM100 There are 4 connectors located on the outside case of the HVM100 The Serial Port input connector external power connector and AD DC Output connector External Power Input Connector Serial Port To...

Страница 96: ...ng with the HVM100 The serial port also provides the printer connec tion Pinout The pinout is as follows 8 4 5 6 2 3 1 7 Pin Num ber Description 1 RTS Request To Send 2 CTS Clear To Send 3 TXD Transmi...

Страница 97: ...ed with the HVM100 is the CBL006 This is a serial connection cable and can be used when com municating through software or printing to a Hyperterminal connection 8 N C No Connection The shell of the c...

Страница 98: ...ucer connection is the input connection into the HVM100 It is located on the top of the HVM100 case The connector used is a standard 4 pin LEMO connector Pinout The pinout is a follows 3 4 1 2 Pin Des...

Страница 99: ...The external power connector is on the top of the HVM100 Pinout The connector pinout is as follows Larson Davis Adapter The power supply normally used with the HVM100 is the Larson Davis PSA027 This...

Страница 100: ...time If the averaging time is set to slow the signal on the DC output pin for the selected channel will be updated once per second If the averaging time is set to 60 seconds the signal on the DC outp...

Страница 101: ...s as follows The signal that is output on each pin is selectable for each channel The selections are the same for each channel We will list the X channel selections however the available selections wi...

Страница 102: ...9 8 HVM100 User Manual 4 11 03...

Страница 103: ...Ammendment 1 1999 E Additionally the current ISO 8041 standard and therefore the HVM100 is compatible with the standards listed below These standards define methods for the measurement of whole body a...

Страница 104: ...30 to 90 RH range This test is performed at 40 C with an input signal of 79 6 Hz Limits of Temperature and Humidity Permanent damage can occur when stored or operated above 60 C or below 20 C Condensa...

Страница 105: ...Clock Calendar Accuracy 0 02 10 C to 50 C 24 hour clock hh mm Year 2000 Compliant Run time Clock One second resolution Format hh mm ss Maximum run time 99 59 59 Power Supply Operates with 2 AA alkalin...

Страница 106: ...1992 Electromagnetic compatibility Generic emission standard Part 1 Residential commer cial and light industry EN 50082 2 1995 Electromagnetic compatibility Generic immunity standard Part 2 Industria...

Страница 107: ...4 11 03 A 5 Direct Input ICP Input Charge Input 50 Signal Generator Ground X Y Z 50 Signal Generator Ground 1 65 k 3300 F X Y Z 50 Signal Generator Ground X Y Z 1000pF 1000pF 1000pF...

Страница 108: ...erence Calibration Frequency Mode Data Buffer Measurements Time History Buffer Measurements Vibration Arms Amin Amax Amp Peak Aeq Arms with optional Peak Hand Arm Arms Amin Amax Amp Peak Aeq A 1 A 2 A...

Страница 109: ...0 4 Hz to 100 Hz Hand Arm Wh 79 6 Hz Whole Body WB Wb Wc Wd We Wg Wj Wk 7 96 Hz Operating Mode Frequency Weighting Reference Calibration Frequency Freq Hz Nominal Freq Hz True Fa dB Tolerance dB 0 10...

Страница 110: ...01 1 1 2 50 2 512 0 00 1 1 3 15 3 162 0 00 1 1 4 00 3 981 0 00 1 1 5 00 5 012 0 00 1 1 6 30 6 310 0 00 1 1 8 00 7 943 0 00 0 10 0 10 00 0 00 1 1 12 5 12 59 0 00 1 1 16 0 15 85 0 00 1 1 20 0 19 95 0 01...

Страница 111: ...z True Fa dB Tolerance dB Freq Hz Nominal Freq Hz True Fb dB Tolerance dB 0 100 0 1000 24 10 2 0 125 0 1259 20 12 2 0 160 0 1585 16 19 2 0 200 0 1995 12 34 2 0 250 0 2512 8 71 2 2 0 315 0 3162 5 51 1...

Страница 112: ...00 1 1 20 0 19 95 0 00 1 1 25 0 25 12 0 00 1 1 31 5 31 62 0 00 1 1 40 0 39 81 0 00 1 1 50 0 50 12 0 00 1 1 63 0 63 10 0 00 1 1 80 0 79 43 0 00 0 100 100 0 0 00 1 1 125 125 9 0 00 1 1 160 158 5 0 00 1...

Страница 113: ...00 2 Freq Hz Nominal Freq Hz True Fb dB Tolerance dB Freq Hz Nominal Freq Hz True Fc dB Wh dB Tolerance dB Ws dB Tolerance dB 0 800 0 7943 36 00 36 00 2 76 00 4 1 00 1 000 32 00 31 99 2 68 00 4 1 25...

Страница 114: ...79 43 0 00 13 88 0 0 00 0 100 100 0 0 00 15 91 1 1 0 00 1 1 125 125 9 0 00 17 93 1 1 0 00 1 1 160 158 5 0 00 19 94 1 1 0 00 1 1 200 199 5 0 00 21 95 1 1 0 00 1 1 250 251 2 0 01 23 96 1 1 0 00 1 1 315...

Страница 115: ...e dB Ws dB Tolerance dB Freq Hz Nominal Freq Hz True Wb dB Wc dB Wd dB Tolerance dB 0 100 0 100 32 04 24 10 24 09 2 0 125 0 1259 28 20 20 12 20 12 2 0 160 0 1585 23 98 16 19 16 18 2 0 200 0 1995 20 23...

Страница 116: ...20 0 19 95 3 00 7 75 19 97 1 1 25 0 25 12 4 44 9 80 21 98 1 1 31 5 31 62 6 16 11 87 24 01 1 1 40 0 39 81 8 11 13 97 26 08 1 1 50 0 50 12 10 09 16 15 28 24 1 1 63 0 63 10 12 43 18 55 30 62 1 1 80 0 79...

Страница 117: ...9 11 1 1 0 500 0 5012 1 27 7 58 7 56 1 1 0 630 0 6310 0 55 6 77 6 77 1 1 0 800 0 7943 0 52 6 42 6 44 1 1 1 00 1 00 1 11 6 30 6 33 1 1 1 25 1 259 2 29 6 28 6 29 1 1 1 60 1 585 3 91 6 32 6 13 1 1 2 00 1...

Страница 118: ...17 47 1 1 100 100 0 43 01 3 01 21 04 1 1 125 125 9 47 46 5 45 25 50 2 2 160 158 5 52 64 8 64 30 69 2 2 200 199 5 58 27 12 26 36 32 2 250 251 2 64 11 16 11 42 16 2 315 316 2 70 04 20 04 48 10 2 400 39...

Страница 119: ...0 19 95 8 2 1 1 25 0 25 12 10 1 1 1 31 5 31 62 12 1 1 1 40 0 39 81 14 2 1 1 50 0 50 12 16 3 1 1 63 0 63 10 18 7 1 1 80 0 79 43 21 6 1 1 Freq Hz Nominal Freq Hz True Wg dB Tolerance dB Freq Hz Nominal...

Страница 120: ...1 2 50 2 512 0 82 1 1 3 15 3 162 1 19 1 1 4 00 3 981 1 74 1 1 5 00 5 012 2 50 1 1 6 30 6 310 3 49 1 1 8 00 7 943 4 70 0 10 0 10 0 6 12 1 1 12 5 12 59 7 71 1 1 16 0 15 85 9 44 1 1 20 0 19 95 11 25 1 1...

Страница 121: ...nge dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V 0 dB See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137...

Страница 122: ...ge 1000pF ICP Gain Noise Floor dB V RMS Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V 0 dB 64 60 dB 74 134 38 dB 99 137 64...

Страница 123: ...Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V 0 dB See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137 20 dB See...

Страница 124: ...dB V RMS Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V 0 dB See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137...

Страница 125: ...Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V 0 dB See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137 20 dB Se...

Страница 126: ...or dB V RMS Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V 0 dB See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 1...

Страница 127: ...oor dB V RMS Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V 0 dB See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99...

Страница 128: ...B V Noise Floor dB V RMS Range dB V Peak Range dB V Noise Floor dB V RMS Range dB V Peak Range dB V 0 dB See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 134 38 dB 99 137 See Note 1 60 dB 74 1...

Страница 129: ...forms to arrive at the results reported by the instrument Description Equation RMS Acceleration T Integration time in seconds aw t instantaneous acceleration t Time in seconds The Aeq integration time...

Страница 130: ...e in seconds to Observation time The linear Arms integration time is controlled by the Averaging time setting a new linear Arms value is calculated and displayed at the end of each inte gration period...

Страница 131: ...ximum reading of all Arms readings from Run to Reset The display is updated at the end of each Averaging time period Minimum Transient Vibration Value Amin minimum reading of all Arms readings from Ru...

Страница 132: ...ctor The CF measurement period is controlled by the Averaging time setting a new CF value is calculated and displayed at the end of each Averaging time period CF is not calculated if the Averaging tim...

Страница 133: ...t 20 commands S1 S19 and Q1 Q19 are reserved for system parameters System parameter settings remain the same regardless of the operating mode The remaining commands S20 S99 and Q20 Q99 are for mode sp...

Страница 134: ...ich there is a specific set of choices the HVM100 returns the index number i e 0 1 2 etc For alphanumeric parameters the HVM100 returns an ASCII string System Parameters Q s and S s The following para...

Страница 135: ...immediately after the OK response from the HVM100 Command Parameter Setting Command Parameter Vibration Hand Arm Whole Body S20 i Operating Mode i 0 to 2 0 Vibration 1 Hand Arm 2 Whole Body i 0 to 2 0...

Страница 136: ...S24 i Gain Z i 0 to 3 0 0 1 20 2 40 3 60 i 0 to 3 0 0 1 20 2 40 3 60 i 0 to 3 0 0 1 20 2 40 3 60 S25 Reserved S26 Reserved S27 Reserved S28 hh Store Time Hour hh 00 to 99 hh 00 to 99 hh 00 to 99 S29...

Страница 137: ...tored as tenths i e 23 2 3 nn 00 to 99 Stored as tenths i e 23 2 3 S35 nn Sum Factor Ky nn 00 to 99 Stored as tenths i e 23 2 3 nn 00 to 99 Stored as tenths i e 23 2 3 nn 00 to 99 Stored as tenths i e...

Страница 138: ...i 5 to 11 5 WB 6 Wb 7 Wc 8 Wd 9 We 10 Wg 11 Wj 12 Wk S40 i AC DC Output X i 0 to 9 0 AC Weighted 1 AC Bandlimit 2 DC rms 3 DC min 4 DC max 5 DC peak 6 DC rms 7 DC min 8 DC max 9 DC peak i 0 to 9 0 AC...

Страница 139: ...imit 2 DC rms 3 DC min 4 DC max 5 DC peak 6 DC rms 7 DC min 8 DC max 9 DC peak i 0 to 9 0 AC Weighted 1 AC Bandlimit 2 DC rms 3 DC min 4 DC max 5 DC peak 6 DC rms 7 DC min 8 DC max 9 DC peak i 0 to 9...

Страница 140: ...ts mV g for Direct and ICP pC g for Charge S48 Sensitivity Z Enter command as S48 n nnne nn n nnne nn Sensitivity of accelerometer Units mV g for Direct and ICP pC g for Charge n nnne nn Sensitivity o...

Страница 141: ...The c indi cates which channel to read X Y Z or S The nn indicates which data to read Hint The HVM100 always returns data in decibels referenced to 10 6 m s2 R Command nn Vibration Hand Arm Whole Body...

Страница 142: ...o available The date information H0 H5 refers to the starting date time of the first history record Hint The HVM100 always returns data in decibels referenced to 10 6 m s2 Command Parameter Settings H...

Страница 143: ...Y Z latching overload indicators are all set The 3 least significant bits of each byte are used to commun micate status information as shown below Byte1 xxxxxnnn Byte2 xxxxxnnn Byte3 xxxxxnnn M1 Run...

Страница 144: ...ored in the nth file location Returns NA if the file location is not available i e less than 0 or greater than 99 M8 n File Empty n 0 to 99 HVM100 will respond as to whether or not the file indicated...

Страница 145: ...se Curves The following are typical frequency response curves for the HVM100 Specifications are subject to change without notice Numerical values given are typical Refer to specific calibration or tes...

Страница 146: ...D 2 HVM100 User Manual 4 11 03...

Страница 147: ...4 11 03 D 3...

Страница 148: ...D 4 HVM100 User Manual 4 11 03...

Страница 149: ...4 11 03 D 5...

Страница 150: ...D 6 HVM100 User Manual 4 11 03...

Страница 151: ...4 11 03 D 7...

Страница 152: ...D 8 HVM100 User Manual 4 11 03...

Страница 153: ...4 11 03 D 9...

Страница 154: ...D 10 HVM100 User Manual 4 11 03...

Страница 155: ...4 11 03 D 11...

Страница 156: ...D 12 HVM100 User Manual 4 11 03...

Страница 157: ...4 11 03 D 13...

Страница 158: ...D 14 HVM100 User Manual 4 11 03...

Страница 159: ...e it is used Follow these steps to measure the DC out put sensitivity Step 1 Calibrate the HVM100 for the accelerometer that will be used Step 2 Setup the HVM100 with the following settings a Operatin...

Страница 160: ...nal Step 7 Stop and Reset the HVM100 Reduce the calibra tor s output level to 0 1 g rms If this isn t possi ble change the HVM100 s weighting to Wh Hand Arm mode The Wh weighting will attenu ate the c...

Страница 161: ...ference voltage is the voltage 1g rms If the DC output is set to peak the reference voltage is the voltage 1g peak AC DC Outputs DC Output For all of the Analog AC and Analog DC outputs the output imp...

Страница 162: ...is approximately 5 0 Vrms a difference of 20 dB Therefore in order for the HVM100 s AC output to function properly with each range i e gain settings of 0 20 40 and 60 dB the HVM100 s gain set tings al...

Страница 163: ...B 5 5000 90 7000 0 5 500 9 700 0 05 50 0 9 70 20 dB 0 5 500 9 700 0 05 50 0 9 70 0 005 5 0 09 7 40 dB 0 0 5 50 0 9 70 0 005 5 0 09 7 0 0005 0 5 0 009 0 7 1 mV g 10 mV g 100 mV g Gain RMS Range g Peak...

Страница 164: ...00 3 The data is based on ideal accelerometers the noise floor and upper limit of the accelerometer were not consid ered 1 mV g 10 mV g 100 mV g Gain RMS Range g Peak Range g RMS Range g Peak Range g...

Страница 165: ...0 0 9 70 20 dB 0 5 500 9 700 0 05 50 0 9 70 0 005 5 0 09 7 1 mV g 10 mV g 100 mV g Gain RMS Range g Peak Range g RMS Range g Peak Range g RMS Range g Peak Range g 40 dB 0 06 50 1 1 70 0 006 5 0 11 7 0...

Страница 166: ...es was calculated using the dBuV range data in the HVM100 User Manual 2 The data in the above table was obtained by electrically testing the HVM100 3 The data is based on ideal accelerometers the nois...

Страница 167: ...repair replacement or exchange of the product Extended Labor Warranty In furtherance of LD s commitment to Total Customer Satisfaction LD offers an extended labor warranty of one 1 year on all product...

Страница 168: ...CT PLUS ANY SHIPPING CHARGES THAT LD MAY BE OBLIGATED TO PAY PURSUANT TO PARAGRAPH C ABOVE G NO EXTENSION OF STATUTE OF LIMITATIONS ANY REPAIRS PERFORMED UNDER THIS LIMITED WARRANTY SHALL NOT IN ANY W...

Страница 169: ...IONS NOT COMPLETED BY AUTHORIZED SERVICE PERSONNEL OR BY FAILURE OF PRODUCTS PARTS FROM SUCH ALTERATIONS 4 COSTS OF REPAIRING DAMAGE CAUSED BY POOR OR IMPROPER MAINTENANCE OR UNAUTHORIZED REPAIR 5 COS...

Страница 170: ...F 4 Warranty Customer Satisfaction April 11 2003...

Страница 171: ...1 Connections on the HVM100 9 1 D Data Communications A 3 Data Key 7 7 Data Storage A 2 Date and Time 3 13 DC Output 3 7 E 3 DC Output Calibration E 1 Declaration of Conformity A 4 Digital Display A 3...

Страница 172: ...ng 5 1 Printing to a printer 5 2 Printing to a Windows Hyperterminal 5 3 R Range Key 4 1 Auto Range 4 1 Cal Level 4 3 Calibrate X Y Z 4 3 Gain X Y Z 4 2 Reference Acceleration 4 5 Sensitivity X Y Z 4...

Страница 173: ...ucers Electrical Impedance A 4 Transverse sensitivity 1 6 Type Precision A 1 E 1 Typical Measurement Ranges for General Purpose Vibration Measurements E 7 Typical Measurement Ranges for Hand Arm and W...

Страница 174: ...HVM100 User Manual 4...