15
Safety Precautions for this Product
Do not use any detachable power cord that is not adequately rated for the unit.
Ne pas utliser un fil électrique amovible qui
n’est pas du tension nominale de
l’appareil.
DO NOT load more than 50 lbs. (23 Kg) on the internal floor of the chamber.
Exceeding this limit may damage the perforated floor and its supports. Excessive
weight in the chamber may increase the risk of it overturning. If your application
requires loading more than 50 lbs., contact Labconco’s Product Service
Department at 800-821-5525 or 816-333-8811 for assistance. Do not exceed the
maximum weight limit for the Drawer, Accessory Shelves, Hanging Bar as shown
in the table below.
Item (each)
Maximum Weight
Limit
Hanging Bar
15 lbs. (6.8 kg)
Drawer
10 lbs. (4.5 kg)
Full Shelf (Accessory)
25 lbs. (11.3 kg)
Half Shelf (Accessory)
15 lbs. (6.8 kg)
Never use flammable gases or solvents in the chamber
– the Hot Plate is an
ignition source. Never use an open flame inside the chamber.
Avoid puncturing the Carbon Filters during installation or normal operation. If you
suspect that a Carbon Filter has been damaged, DO NOT use the chamber;
contact Labconco at 800-821-5525 or 816-333-8811 for further information.
Avoid direct exposure of plastic or coated materials to ultraviolet (UV) radiation.
When cleaning the chamber:
o
Always wear appropriate personal protective equipment (PPE).
o
NEVER use high pressure water to clean the liner.
o
NEVER spray cleaners/disinfectants into the blower intake at the top, right
side of the interior liner; this may damage the blower.
o
Avoid splashing any cleaning solution on skin or clothing.
o
Ensure adequate room ventilation & use Cleaning Cycle (see
Section
7:Cleaning Cycle
).
o
Carefully follow any cleaning solution’s safety instructions.
o
Ensure adequate ventilation.
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...