Chamber Operational Checks
Prior to use with evidence, you should perform the following steps and operational tests
to ensure the chamber is operating properly.
Electrical System Check
After plugging the power cord into the top, back of the unit, and into an
appropriate wall outlet, turn the power switch on (located on front of main control
panel
– see Fig. 4-2).
The Display Screen should show the CApture BT logo, and then go to the Main
Menu Screen. You should hear a single beep after turning the power switch on.
The interior chamber light should also come on.
If the above checks are good, the electrical systems are working, and you can
turn the unit back off.
Pre-Run Steps
(may be performed during
Initial Chamber Set-up
– see Section 6)
Verify the power cord is plugged into the top, back of the unit, and into an
appropriate wall power outlet.
Locate the Water Bottle inside the Door labeled “H2O”. Remove the Water Bottle
and fill it with water (RO Filtered or Distilled Water are recommended).
Replace the Water Bottle in the holder and make sure the water line hose goes
back through the hole in the Water Bottle Lid and the end of the hose reaches the
bottom of the Water Bottle.
Turn the power switch on (see Fig. 4-2). You should hear the water pump turn on and
see the water level slowly dropping in the Water Bottle. The pump should run for
approximately 60-90 seconds, and will remove approximately one-third to one-half of
the water from the Water Bottle.
IMPORTANT
Verify the valve behind
the Water Bottle is
flipped to the down
position, which is
NORMAL FILL, as
shown in Fig. 4-3.
Figure 4-3
VALVE
WATER
BOTTLE
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...