Tubing Cleanout (As Needed)
Over time the internal air paths within the tubing behind the Right Side Service Panel
may accumulate CA residue/particulate. The CA typically adheres to the internal walls
of the tubing that sends air from the Recirculation Blower to the CA Chamber (Hot Plate
area). The CA build-up is seen as a white, fine powder. If this build up is not cleaned
out, airflow may reduce and the white CA particulate can begin to circulate/deposit in
the Main Chamber. To remove CA build-up in the tubing, follow these steps:
1. If any evidence is present inside the Main Chamber, remove it. If a glue tin is
present on the Hot Plate, remove it.
2. Obtain a shopvac or similar vacuum with HEPA-Filter.
3. Open the CA Chamber (Hot Plate) Door.
4. Insert the end of the vacuum hose approximately 3-6 inches (7-15 cm) inside the
tube opening behind the Hot Plate. See Fig. 7-10.
HOT PLATE
TUBE
OPENING
Figure 7-10