46
Humidify
– Program Set-up
If re-humidifying prints, you can utilize the Humidify Cycle functionality to raise the
chamber
’s relative humidity to any user-selectable level, up to 80% maximum, and hold
it at that level for a user-selectable time period.
On the Main Menu screen, Select
RUN
by moving the red arrow to Run menu option,
and press
[OK]
. The Run Menu (see Fig. 6-19) displays three sub-menus, which offer
three types of cycles. Select
HUMIDIFY
by moving the red arrow to Humidify menu
option, and press
[OK]
. The Humidity Menu (see Fig. 6-19) will be displayed:
The Set Point values for
Humidity
and
RH Dwell
can be adjusted to desired levels. The
selected value will blink, and you can move between the Humidity Set Point value, and
the RH Dwell Set Point value with
[UP/DOWN]
. Use
[RIGHT/LEFT]
to increase or
decrease the selected value. When the values are set to user preference, press
[START]
to begin the Humidify Program. The RH Dwell time will not begin until the
target RH level is achieved. While the Humidify Cycle is running, the following screen
will be displayed (Fig. 6-20):
Figure 6-19
CURRENT
RELATIVE
HUMIDITY
TIME SINCE RH
DWELL PERIOD
BEGAN
TARGET
RELATIVE
HUMIDITY LEVEL
TOTAL TIME TO
DWELL AFTER
TARGET RH
ACHIEVED
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...