The Program Parameters (shown in Fig. 6-8) can be adjusted to desired levels within
certain minimum and maximum values. Programs 1-3 are factory defaults for running
common fuming cycles. The parameters can be modified in the first three programs,
and a cycle ran; however, the modifications will not be saved, and will revert back to the
factory default parameters. Programs 4-20 will save modifications to the program
parameters. The modifications will be saved once a cycle has begun.
To browse through the programs, move the red arrow to
PROGRAM #
using
[UP/DOWN]
. Then, use
[RIGHT/LEFT]
to change to the desired Program #.
The CA Fume Parameters can be selected by using
[UP/DOWN]
to highlight the
desired parameter, then use
[RIGHT/LEFT]
to change the selected parameter value.
The parameters are described below:
Grams CA
Min
– 0.0
Max
– 99.9
This parameter is for user reference only, and will be displayed on the following screen
just before the Fuming Cycle is initiated. Reminds user how much CA to place in the tin
to achieve consistent development results.
Humidity
Min
– 20 %
Max
– 80 %
This parameter sets the desired level of relative humidity the chamber will achieve
before entering the Fuming stage of the program. Adding humidity to a desired level
rehydrates prints, which allows for better CA polymerization, which improves print
development. Set in 1% increments.
RH Dwell
Min
– 0:00
Max
– 99:30
This parameter sets the desired dwell time between when the desired relative humidity
level is reached in the chamber, and when the Fuming portion of the program begins. If
set to 0:00 the Fuming portion will begin immediately after the desired relative humidity
level is reached. Set in 30 second increments.
CA Heat
Min
– 100
O
F (38
O
C)
Max
– 425
O
F (218
O
C)
This parameter sets the temperature of the hot plate upon which the tin of CA will be
placed. CA volatizes quickly around 170
O
F (77
O
C); however, some users prefer higher
temperatures to expedite volatizing of CA fumes. Set in 1
O
C (~2
O
F) increments.
Fume
Min
– 0:00
Max
– 99:30
This parameter sets the duration of the Fuming stage. Time begins on this stage when
the Hot Plate turns on, not when the desired temperature is reached. Set in 30 second
increments.
Purge
Min
– 4:00
Max
– 10:00
This parameter sets the duration of the Purge, or Exhaust, stage of the program. Under
ideal conditions with nominal CA glue in the glue tin, chamber purges completely in 4
min. Set in 30 second increments.
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...