72
two (2) nuts on Limit Switch Bracket. Lift Limit Switch and Bracket over threaded studs.
Remove two (2) screws & nuts holding Limit Switch to Bracket.
Before assembling
new Limit Switch, note that the old Limit Switch did not have the silver, metal
tang (tab), remove the silver, metal tang (tab) off of the new Limit Switch by
twisting it with pliers.
22. CA Door Limit Switch
– P/N 3832300
Note wire locations on switch for correct reassembly, then disconnect wiring. Open CA
Chamber Door. Remove the two (2) nuts on top of the Limit Switch to remove the Limit
Switch.
23. Humidifier Fan
– P/N 3377901
Disconnect two-pin wire harness connector. Remove four (4) screws and lift fan off of
Humidifier Cover.
24. Water Bottle
– P/N 3175500
Remove Water Bottle from cradle, pull water line out of lid hole.
25. Water Filter
– P/N 3186800
Pull water hose off each end. Twist Filter Housing open to clean filter, or replace with
new Filter.
26. Toggle Valve
– P/N 3185500
Take careful notes of each water line tube and which port it is connected to (ports are
numbered on valve body), or take several pictures. Remove water line tubing from all
five (5) ports. Loosen and remove the jam nut and lockwasher located by the Toggle on
the INSIDE of the Water Bottle Compartment. The threaded hose barb fittings from the
old Toggle Valve will need to be reused, or new can be ordered (P/N 4595108). Use
Teflon tape on hose barb threads.
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...