Humidify Timeout
During the Humidify stage within a Program, if the user-selected humidity is not reached
within a certain time, an error occurs, and the program ends prematurely. The screen in
Fig. 7-22 is displayed. This protects the unit, and alerts the user a problem may exist.
However, if the Fuming Chamber is used in an environment with very low humidity, the
unit may require more time to reach the desired level of humidity. Before changing the
default Humidify Timeout (35 min.), make sure no other problems exist within the unit.
Contact Labconco Service Department at (800) 821-5525 or +1 (816) 333-8811 for
troubleshooting assistance.
To adjust the Humidify Timeout, following the procedure below:
1. On the
MAIN
Menu screen, select
SERVICE
, press
[OK]
. Enter the Service
Password:
[UP] [DOWN] [LEFT] [RIGHT] [OK]
.
Select
RH TIMEOUT
on the
SERVICE
Menu and press
[OK]
. The following screen will be displayed (Fig. 7-
23).
2. Press the
[RIGHT]
button to increase the timeout value in 5 minute increments,
press
[OK]
.
Figure 7-22
Figure 7-23
PRESS
[RIGHT]
OR
[LEFT]
TO
ADJUST
TIMEOUT VALUE
(IN MINUTES)
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...