9
2: Before You Install
Before you install the product, the site should be prepared for installation. Examine the
location where you intend to install it. You must be certain that the area is level and of
solid construction. In addition, a dedicated source of electrical power must be located
within 10 feet (3 m) of the installation site.
Location Requirements
The chamber is equipped with leg levelers (if not attached to the accessory stand) to
ease level chamber installation. It should be located on a flat surface for proper
chamber alignment and door operation. The surface must support 500 lbs. (227 kg).
Clearance Requirements
A minimum clearance of at least 6 inches (150 mm) is suggested on the top of the
chamber, 12 inches (300 mm) on the right side for service, and 2 inches (50 mm)
behind the chamber. There should be 36 inches (914 mm) clearance at the front of the
chamber to allow the door to swing open without hitting any obstructions.
See
Appendix B: Dimensions
for overall product dimensions.
Electrical Requirements
Catalog Number
Typical Operating
Current (Amps)
Electrical Circuit Requirements
1
3170000
4 A
115 V, 60 Hz, 12 A
1 Phase
3170001, -02, -03, -04, -05
2 A
230 V, 50/60 Hz, 6 A
1 Phase
An outlet with an appropriate circuit breaker should be located as close as possible to
the product, but no greater than 10 feet (3 m). Consult your local electrical codes for
properly rated circuit breakers. For safe operation the dedicated outlet must provide a
protective earthing ground connection to the product.
1 Ele
ctrical Requirements, ‘V’ = VAC (Voltage with alternating current), ‘A’ = Amperes
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...