Drain Humidifier Tank (Weekly)
Replace water in Humidifier Basin and Water Bottle. This is important to do weekly,
especially before the chamber will sit unused for several days (typically on a Friday
before sitting unused over the weekend). Follow the steps below to empty the
Humidifier Basin and Water Bottle.
1. Open the Water Bottle Door, locate the Valve Toggle (see Fig. 7-1) on the back
wall, just to the left of the Water Bottle. Next, flip the Valve Toggle “up” which
reverses water flow back into the Water Bottle.
2. Make sure the unit is plugged in, and Main Power Switch is ON.
3. Locate the Momentary Switch (see Fig. 7-1) on the top panel inside Water Bottle
Compartment. Press and hold this switch for approximately 15 seconds. When
the switch is pressed, you should hear the Water Pump turn on, and see the
water level in the Bottle rising. Release the Switch, Water Pump should continue
to run on its own, if not, hold Switch for another 5 seconds.
4. When air bubbles are seen coming from hose inside Water Bottle, turn the Main
Power Switch OFF. Empty the Water Bottle of all water. Flip the Valve Toggle
back “down” to the NORMAL FILL position. Replace Water Bottle. Before using
chamber again, fill Water Bottle with water.
MOMENTARY
SWITCH
VALVE
TOGGLE
WATER
BOTTLE
Figure 7-1
Содержание Capture BT 31700-00
Страница 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Страница 55: ...55 Figure 7 6...
Страница 81: ...81 Wiring Diagrams 100 120V...
Страница 82: ...82 208 230V...
Страница 87: ...87 Chamber will not achieve humidity...
Страница 88: ...88 Chamber does not exhaust CA fumes...
Страница 89: ...89 Door alarm active when door s closed...
Страница 90: ...90 Hot Plate will not achieve temperature set point...
Страница 91: ...91 White CA residue builds on inside of Chamber...