18
If using the MRS-101A-V (1ch-Voltage Transmitter)
•The product does not have a sensor checking function for confirming the sensor connection.
Do not start measuring until you have confirmed that the input is connected.
•If the “CHECK” switch is pressed, the receiver operates differently depending on the firmware version as
shown below.
Firmware version
"CHECK" LED
"DATA" LED
Receiver Analog Output
Ver.1.00
Lit in green
●
↓
Off
●
Lit in green
●
(Lit for three seconds)
For approximately five
seconds
Approximately 2.5 V
output
Ver.2.00 or later
Off
●
Off
●
No change
l)
"START" switch, "START" LED
The "START" switch is for starting measurement. The "START" LED below the "START" switch lights up in
green.
When the "START" switch is pressed, the product transitions from Command Mode to Measuring Mode and starts
outputting measurement data from the receiver’s analog output.
The "DATA" LED lights up in red for three seconds if a wireless communication error occurs even in a single
packet.
Measurement can only be started in Command Mode, and after the product switches to Measuring Mode, no
switches other than "STOP" and "MUTE" are available.
• When measurement starts normally:
1) When the "START" switch is pressed, the "DATA" LED, "LOCK" LED, and "START" LED light up in
green.
2) The product transitions from Command Mode to Measuring Mode.
3) One to three "Transmitter’s battery voltage checking" LEDs and one to three "RSSI" LEDs light up in green.
4) Outputting of measurement data begins from the receiver’s analog output.
• When measurement fails to start normally (due to a wireless communication error, etc.):
1) When the "START" switch is pressed, the "DATA" LED lights up in red for three seconds, and the buzzer
sounds.
(The buzzer does not sound if "MUTE" is off.)
2) Measuring Mode does not begin, and the product remains in standby in Command Mode.
"START" LED
"DATA" LED
Status
Lit in green
●
Lit in green
●
Switched to Measuring Mode. Measurement has started.
Begin outputting measurement data from the receiver’s analog output.
Off
●
Lit in red
●
(Lit for three
seconds)
Cannot transition to Measuring Mode. The product remains in standby
in Command Mode.
Try again in an environment that enables normal wireless
communication.
In
approximately
Seven seconds
Содержание MRS-100 Series
Страница 74: ...73 8 7 OUTSIDE DRAWING 8 7 1 MRS 114A 4ch Receiver ...
Страница 75: ...74 8 7 2 MRS 104A S 4 ch Strain Transmtter ...
Страница 76: ...75 8 7 3 MRS 101A S 1ch Strain Transmitter ...
Страница 77: ...76 8 7 4 MRS 101A SE 1ch Strain Transmitter EXT Antenna ...
Страница 78: ...77 8 7 5 MRS 101A V 1ch Voltage Transmitter ...
Страница 79: ...78 8 7 6 DA DB 05RP SMA 08 High gain dipole antenna ...
Страница 81: ...Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...